Z-0021-2025Class II

The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque...

Stryker Corp. / initiated 2024-08-30 / Open, Classified / root cause: labeling and UDI

Stryker Neurovascular began a recall of The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the... on . FDA classified it as Class II and gives the reason as "Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0021-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Stryker Corp.
Product
The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. A rotating hemostatic valve with side-arm adapter is provided with each microcatheter.
Product code
QJP Catheter, Percutaneous, Neurovasculature
Reason for recall
Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to customers in the US.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On 09/12/2024, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL - Customer Notification" Letter to customers informing them that certain lots of Trevo Trak 21 Microcatheters were shipped with the incorrect configuration. The affected products contained Instructions for Use (IFUs) and product labels for international use, instead of the United States (US). The International IFUs contain intended uses that have not been cleared/approved for the U.S. market. Customer are instructed to: 1. Immediately check internal inventory for affected devices. 2. When using impacted lot(s) of Trevo Trak 21 Microcatheters, follow the US IFU attached in this communication NV00055463-02 (Attachment A) instead of the one that originally came with the device. 3. Circulate recall notification internally to all interested/affected parties and post this customer notification on or near affected products. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. Inform Stryker Neurovascular if the subject devices have been distributed to other organizations and provide contact details so that Stryker Neurovascular can inform the recipients appropriately. 6. Inform Stryker Neurovascular of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916- 3876), Email: [email protected]). 7. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form allow for updating records and will also negate the need to send any further unnecessary communications on this matter. 8. Email the completed form to [email protected]. Contact the local Stryker representative, send an email to [email protected] or call US Customer Service: 1-855-91-NEURO (1-855-916-3876) for further assistance related to this Medical Device Recall Customer Notification.
Quantity in commerce
6
Distribution
U.S. Nationwide distribution in the states of CA and VA.
Product codes and lots
Model/Catalog Number: 90338 UDI Codes: (01)07613327313956(17)251125(10)0000590444 (01)07613327313956(17)260101(10)0000603275 Lot Numbers: 0000590444 0000603275

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0021-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0021-2025Class IIOngoingVoluntary: Firm initiated