Melrose Wakefield Hospital REF K05T-2159 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The...
Merit Medical Systems, Inc. began a recall of Melrose Wakefield Hospital REF K05T-2159 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and... on . FDA classified it as Class II and gives the reason as "Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0062-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Melrose Wakefield Hospital REF K05T-2159 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pressure of that fluid.
- Product code
- DXT Injector And Syringe, Angiographic
- Reason for recall
- Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 08/30/2024, the firm email and/or UPS overnight delivery an "URGENT MEDICAL DEVICE RECALL NOTICE" informing customers that specific lots of MONARCH Inflation Devices due to a packaging defect, there may be a small hole in the Tyvek sterile barrier of the device packaging. Use of the affected product may result in exposure to pathogens/pyrogens which could lead to infection. Customer are instructed to: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan, and email the completed CRF to Customer Service at [email protected] within 7 calendar days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. For questions - contact your Merit Sales Representative or Merit Customer Service via email at [email protected] or via phone at +1 800 356 3748 | Hours: 6 am to 6 pm MST | Mon-Fri.
- Quantity in commerce
- 432 devices
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Belgium, Canada, France, Japan, Monaco, Netherlands, Qatar, United Arab Emirates, United Kingdom.
- Product codes and lots
- Catalog Number: K05T-02159 UDI-DI code: 00884450140840 Lot Number: T2417859 T2434208 T2509792 T2549931 T2561788 T2625814 T2675576 T2762716 T2801089 T2893877
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0057-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0058-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0059-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0060-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0061-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0062-2025 | Class II | Ongoing | Voluntary: Firm initiated |