Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0051-2025Class II

Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF...

Auris Health, Inc / initiated 2024-08-02 / Open, Classified / root cause: design

AURIS HEALTH INC began a recall of Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, MON-000006-RFB on . FDA classified it as Class II and gives the reason as "Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0051-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Auris Health, Inc
Product
Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, MON-000006-RFB.
Product code
EOQ Bronchoscope (Flexible Or Rigid)
Reason for recall
Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
Root cause tag
design
FDA root cause
Device Design
Action
On 8/2/24, field safety notices were emailed to customers informing them of the following: 1) If the flow of electricity stops, the system may switch to battery backup power to allow for safe system shutdown. 2) Prior to use, carefully inspect the cords for any signs of excessive wear or damage while not plugged into an electrical outlet. 3) Firm is sourcing new replacement power cords to increase their robustness. New power cords will be provided as replacements once they are available. A member of MONARCH support will contact you to coordinate the replacement. 4) Forward this notice to any personnel in your facility who need to be informed. If any affected products have been forwarded to another facility, contact that facility and provide a copy of this notice to the relevant personnel. 5) Complete are return the acknowledgement form via email to [email protected] 6) If you feel that your power cord is damaged or defective or have questions , please contact MONARCH Support at 1-800-434-0032 or [email protected] to obtain a replacement that has been screened for this defect.
Quantity in commerce
176
Distribution
Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.
Product codes and lots
Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00. REF/UDI-DI/Serial Numbers: MON-000005-01/+B634MON000005010/110024, 13, 110014, 110016, 110040, 110038, 110027, 110035, 110022, 110034, 1806P2, 120170, 110045, 110039, 110049, 110047, 110052, 110030, 110032, 110021, 110043, 110050, 110029, 110033, 110046, 110048, 110018, 110015, 110051, 1806P1, 110012, 14, 110041, 110036; MON-000005-01R/+B634MON00000501R0/110020, 110017, 110031, 110028, 120042, 110013. MON-000006/+B634MON0000060/120088, 120057, 120046, 120058, 120121, 120152, 120015, 120160, 120068, 120167, 120025, 120030, 120044, 120093, 120040, 120063, 120092, 120173, 120036, 120064, 120122, 120056, 120128, 120017, 120089, 120075, 120031, 120169, 120113, 120168, 120055, 120148, 120014, 120060, 120124, 120164, 120123, 120007, 120006, 120004, 120120, 120126, 120082, 120047, 120051, 120020, 120103, 120153, 120108, 120037, 120074, 120005, 120130, 120095, 120096, 120009, 120111, 120022, 120054, 120163, 120027, 120090, 120141, 120072, 120155, 120101, 120143, 120125, 120052, 120100, 120050, 120098, 120045, 120131, 120013, 120105, 120165, 120136, 120029, 120041, 120139, 120137, 120053, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 120145, 120127, 120011, 120070, 120076, 120134, 120117, 120077, 120149, 120102, 120002, 120010, 120003, 120023, 120146, 120043, 120071, 120107, 120065, 120116, 120150, 120066, 120166, 120033, 120079, 120008, 120132, 120086, 120032, 120073, 120016, 120080, 120078, 120012, 120084, 120083, 120038, 120067, 120021; MON-000006-RFB/+B634MON0000060/120081, 120026, 120091, 120106, 120104.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0051-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0051-2025Class IIOngoingVoluntary: Firm initiated
Z-0052-2025Class IIOngoingVoluntary: Firm initiated