Monarch Platform (Urology), REF: MON-000007, MON-000007-RFB
AURIS HEALTH INC began a recall of Monarch Platform (Urology), REF: MON-000007, MON-000007-RFB on . FDA classified it as Class II and gives the reason as "Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0052-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Auris Health, Inc
- Product
- Monarch Platform (Urology), REF: MON-000007, MON-000007-RFB
- Product code
- FGB Ureteroscope And Accessories, Flexible/Rigid
- Reason for recall
- Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 8/2/24, field safety notices were emailed to customers informing them of the following: 1) If the flow of electricity stops, the system may switch to battery backup power to allow for safe system shutdown. 2) Prior to use, carefully inspect the cords for any signs of excessive wear or damage while not plugged into an electrical outlet. 3) Firm is sourcing new replacement power cords to increase their robustness. New power cords will be provided as replacements once they are available. A member of MONARCH support will contact you to coordinate the replacement. 4) Forward this notice to any personnel in your facility who need to be informed. If any affected products have been forwarded to another facility, contact that facility and provide a copy of this notice to the relevant personnel. 5) Complete are return the acknowledgement form via email to [email protected] 6) If you feel that your power cord is damaged or defective or have questions , please contact MONARCH Support at 1-800-434-0032 or [email protected] to obtain a replacement that has been screened for this defect.
- Quantity in commerce
- 4
- Distribution
- Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.
- Product codes and lots
- Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00. REF/UDI-DI/Serial Numbers: MON-000007/+B634MON0000070/120059, 120133, 210002; MON-000007-RFB/10810068810889/210001.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0051-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0052-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.