Z-0053-2025Class II

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2024-08-28 / Open, Classified / root cause: packaging

MEDLINE INDUSTRIES, LP - Northfield began a recall of Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket on . FDA classified it as Class II and gives the reason as "CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0053-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket
Product code
FDE Laparoscopy Kit
Reason for recall
CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.
Root cause tag
packaging
FDA root cause
Unknown/Undetermined by firm
Action
The recalling firm issued letters on and dated 8/28/2024 via first class mail and email. The letter explained the reason for recall and instructed the customer that nothing was to be returned to Medline. The customer is to request a sticker to be placed on affected kits instructing the user to remove the recalled component. The consignee is required to immediately check their stock for the affected item number and affected lot numbers which can be located within the recall portal and quarantine all affected product immediately. The consignee is to use the link and information in the letter to complete their response form. They are to list the quantity of affected product in inventory on the form and submit the form. Upon receipt of the submitted response form, the consignee will receive over-labels to place on affected inventory with instructions for the user to remove the affected component prior to using the kit. The labels will be provided via FedEx overnight. If the consignee is a distributor or has resold or transferred the product to another company or individual, the consignee is to notify their downstream customers of this communication.
Quantity in commerce
72 kits
Distribution
US Nationwide distribution in the state of CA.
Product codes and lots
Lot numbers 24AMA209, exp. 2025-06-30; 24BMB025, exp. 2025-08-31; 23LMA745, exp. 2025-07-31; UDI-DI each 10195327384494, UDI-DI case-40195327384495

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0053-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0053-2025Class IIOngoingVoluntary: Firm initiated