Z-0230-2025Class II

LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o...

Physio Control, Inc. / initiated 2024-08-30 / Open, Classified / root cause: supplier

Physio-Control, Inc. began a recall of LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o... on . FDA classified it as Class II and gives the reason as "Due to an out of tolerance tool being used on monitor/defibrillator systems". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-0230-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Physio Control, Inc.
Product
LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99577-001955 o 99577-001956 o 99577-001957 o 99577-001958 o 99577-001962 LIFEPAK 15 intended use monitor/defibrillator.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Due to an out of tolerance tool being used on monitor/defibrillator systems.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
On 09/30/2024, the firm mailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that Stryker discovered that an out of tolerance service tool was used on affected LP15 devices in which there is a potential for Noninvasive Blood Pressure (NIBP), System or Auxiliary Power Connectors to be loose. There is a possibility that water or dust may accumulate in the ingress and can lead to permanent or intermittent failures. Customers are instructed to: 1. Immediately check internal inventory to locate the product listed on the attached business reply form. 2. To confirm receipt of this Medical Device Correction Notice and understanding of the provided information, email the enclosed Business Reply Form (BRF) by October 31st, 2024 date, to [email protected]. 3. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for corrective action on the device. 4. There are no additional actions that users can and should take. Please continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter Maintaining the Equipment subsection General Maintenance and Testing . 5. In the interim, Stryker recommends that you continue to use your LP15 device. 6. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until the correction has been completed. 7. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: [email protected] b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/ If you have any questions or concerns, please contact Customer Service +1 800 442 1142 or by email [email protected].
Quantity in commerce
105 systems
Distribution
U.S.: CT, GA, MA, NY, OH, WA, and WV. O.U.S.: N/A
Product codes and lots
LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99577-001955 o 99577-001956 o 99577-001957 o 99577-001958 o 99577-001962 UDI-DI codes: o 00883873808214 o 00883873988022 o 00883873979426 o 00883873979358 o 00883873911693 o 00883873979198 o 00883873911648 o 00883873884386 o 00883873871713 o 00883873871706 o 00883873871690 o 00883873871683 o 00883873911631 o 00883873871645 Serial Numbers: o 45435833 o 46470647 o 46686048 o 46686573 o 46686708 o 46687360 o 46906073 o 49037777 o 49704552 o 38166647 o 38282988 o 38290136 o 38291331 o 38802782 o 39541663 o 39541666 o 39752193 o 40259639 o 41446867 o 41447020 o 41453331 o 42325059 o 43574409 o 43984215 o 45040678 o 45308104 o 45455023 o 46017745 o 46572054 o 46576300 o 48349664 o 48605645 o 48605663 o 49001744 o 49025185 o 49070762 o 49087040 o 49087053 o 49224619 o 49374174 o 49422314 o 49619921 o 38200440 o 38805564 o 38805583 o 39382735 o 41446693 o 41446882 o 41446975 o 41446977 o 41447027 o 41453239 o 41893891 o 42324176 o 42610924 o 43181935 o 43182268 o 43182389 o 43462227 o 43465372 o 43487806 o 43610792 o 43791346 o 43841374 o 43860850 o 43862247 o 43869747 o 44169862 o 44356379 o 44360003 o 44601246 o 45590467 o 45907317 o 46364496 o 46691551 o 47447195 o 47515791 o 47539203 o 47539310 o 47785570 o 47794250 o 48524850 o 48583111 o 48585709 o 48600317 o 48605407 o 48605467 o 48704914 o 48749443 o 49022830 o 49031342 o 49036187 o 49056308 o 49057928 o 49123367 o 49126651 o 49127012 o 49132399 o 49132740 o 49134638 o 49511402 o 49513814

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0230-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0230-2025Class IIOngoingVoluntary: Firm initiated