Leica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c)...
LEICA BIOSYSTEMS NUSSLOCH GMBH began a recall of Leica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c) 14901850001R0501, d) 1491850UVUL; for freezing and... on . FDA classified it as Class II and gives the reason as "As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0087-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Leica Biosystems Nussloch GmbH
- Product
- Leica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c) 14901850001R0501, d) 1491850UVUL; for freezing and sectioning tissue samples
- Product code
- IDP Microtome, Cryostat
- Reason for recall
- As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Leica Biosystems issued an Urgent Field Action Notice to its consignees on 09/05/2024 via mail. The notice explained the problem with the device and requested the placement of a warning sticker regarding the use of flammable freezing sprays on the device and the addition of a supplemental sheet to the current IFU. For questions, please contact: Heather Jensen, Leica Biosystems, 5205 US Highway 12, Richmond, IL 60071 USA, Tel.: +1-815-298-8371, [email protected]
- Quantity in commerce
- 322 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- a) Material Number 14047131148, UDI/DI 04049188064790, Serial Numbers: 598061998, 1874062001, 329111997, 370121997, 175081997, 517041998, 376021998, 702071998, 2935062002, 333041998, 693071998, 3524062003; b) Material Number 14901850001, UDI/DI 04049188003423, Serial Numbers: 1922112000, 2396082001, 2032122000, 1850092000, 3728, 1885, 3156102002, 3441032003, 1684082001, 3682102003, 3080102002, 4791, 1439, 3517062003, 1186071999, 5742, 3976022004, 217091997, 960021999, 2097022001, 4660052005, 3107, 2805, 3425032003, 1884092000, 729101998, 1407121999, 1678062000, 2870052002, 3346022003, 1716062000, 3222112002, 2144042001, 4211072004, 1542032000, 3069102002, 469021998, 4955032006, 5674052008, 2873052002, 915121998, 2812042002, 1742072000, 4720, 5508, 4947032006, 4137052004, 3318012003, 3357032003, 4548032005, 1110061999, 4722062005, 4721, 5507, 2925, 4129042004, 2628122001, 2509102001, 2136042001, 4118, 3508062003, 3421, 3400022003, 4180062004, 3496052003, 3321012003, 5307032007, 5620032008, 1336, 3322012003, 3980, 3276122002, 3306012003, 979021999, 2068022001, 1887092000, 3207, 1031031999, 2968072002, 2788032002, 4231072004, 794101998, 454051999, 5668, 4527022005, 952021999, 1663, 1387121999, 2172, 3760112003, 2936062002, 1076051999, 3558062003, 2939062002, 2906052002, 4264092004, 4471, 3333022003, 2363092001, 3876022004, 2586122001, 3435042003, 4252092004, 2903052002, 5064, 5755, 1856092000, 777101998, 5361, 5607, 2736022002, 2750022002, 917011999, 3689102003, 3197, 2260062001, 4049042004, 4925022006, 1843092000, 5315042007, 4167, 2911082002, 447031998, 2461102001, 3415, 3603082003, 904, 2145022001, 2028122000, 5660052008, 5052, 2861042002, 1321101999, 453, 3008, 3096102002, 2717022002, 3193112002, 3030082002, 961021999, 3202, 1153061999, 475031998, 2782, 5465, 5579, 3631092003, 5030062006, 3168102002, 3251122002, 5141092006, 3927, 2227052001, 757101998, 573051998, 5267, 2707, 3598, 220042001, 4833102005, 3024082002, 2890052002, 4593032005, 2335072001, 5108092006, 3488, 1878062001, 2757032002, 1063041999, 3679, 1310101999, 1561032000, 3647092003, 4466012005, 2798032002, 4159, 5095082006, 4589032005, 3068102002, 2632012002, 1315, 2215042001, 3869022004, 2425092001, 4160052004, 3354022003, 4774, 5001052006, 5740092008, 3445, 4898, 4954032006, 1164071999, 5355062007, 4976042006, 4728062005, 3407032003, 5727, 4588, 5280, 5659, 5814, 4639052005, 416011998, 3316, 3499052003, 1121061999, 5805112008, 2909062002, 5716072008, 4202, 4816, 4355102004, 2780032002, 4228, 2089022001, 1776082001, 2139, 5335, 4861112005, 608051998, 1753, 2735022002, 2599112001, 5393072007, 5813122008, 5531122007, 332111997, 3025, 4608032005; c) Material Number 14901850001R0501, UDI/DI 04049188003423, Serial Numbers: 4723; d) Material Number 1491850UVUL, UDI/DI 04049188002730, Serial Numbers: 216052005, 482, 765012007, 332, 1067, 1187012008, 1208, 417, 511072006, 1095102007, 1347, 1200, 970082007, 370, 895, 494062006, 454082006, 870052007, 1488, 672122006, 1149122007, 1059, 1345, 658122006, 1196012008, 301092005, 772012007, 708012007, 709012007, 655112006, 802022007, 868052007, 885, 698122006, 513072006, 378, 1453, 991082007, 201052005, 1076102007, 1389, 997, 610, 265082005, 671122006, 896, 1137, 429032006, 1317052008, 949, 987082007, 189, 646112006, 597, 1307042008, 1128112007, 792022007, 771, 576092006, 1272, 1259032008, 1524, 382, 1313052008, 1255032008, 1325, 1017, 1016, 1055102007, 702, 345112005, 1044102007, 1047102007, 1403, 526, 602102006, 455042006, 894, 1320052008
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0085-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0086-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0087-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0088-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0089-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0090-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.