MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATOR, 1.5oz tube (Model...
Integra LifeSciences Corp. began a recall of MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATOR, 1.5oz tube (Model Number 31515) and 3.5oz tube (Model Number 31535) on . FDA classified it as Class II and gives the reason as "There is a potential for pinholes in the applicator pouch film which creates a sterility concern". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0054-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Integra LifeSciences Corporation
- Product
- MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATOR, 1.5oz tube (Model Number 31515) and 3.5oz tube (Model Number 31535)
- Product code
- FRO Dressing, Wound, Drug
- Reason for recall
- There is a potential for pinholes in the applicator pouch film which creates a sterility concern.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- An URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION letter dated 9/13/24 was sent to customers. Actions to be Taken by Customers (Medical Facility): 1. Complete the Medical Facility Acknowledgement Form below. 2. If you do have units of the impacted product (Table 1) remove the applicator immediately from service and discard it. 3. Forward this notice to those who utilize the product so they are aware of this recall and can identify any affected product that may remain in clinical areas. 4. Return the Medical Facility Acknowledgement Form to [email protected] or FAX to 1-609-750-4220. 5. Keep a copy of the form for your records. OR Actions to be Taken by Distributors: 1. Complete the Distributor Acknowledgement Form below. 2. Please print and include a copy of this notice with every shipment of impacted products (Table 1) that have not yet been shipped to customers. 3. Return the Acknowledgement Form to [email protected] or FAX to 1-609-750-4220. 4. Keep a copy of the form for your records. 5. Forward a copy of this notice to any customers that have purchased the impacted products (Table 1). Receipt of this form ensures that Integra has achieved a level of effectiveness in communicating this information. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that our customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected]
- Quantity in commerce
- 120124 units
- Distribution
- Worldwide - US Nationwide distribution in the states of MN, IL, UT, FL, NC, CA, MO, IN, MS, OH, PA, TX, HI, MA, WA, GA, AZ, VA, NY, CO, MI, OK, NV, NJ, SC, OR, VT, DC, WI, TN, WV, KS, AL, NE, MD, ID, LA, AR, KY, IA, AK, NM, ME, RI and the countries of Madagascar, Malaysia, Guam.
- Product codes and lots
- Model 31515 UDI-DI: 10381780486824; Model 31535 UDI-DI: 10381780486831; All lots distributed from 12/17/2020 to 07/22/2024 with expiration dates 08/01/2024 to 06/01/2028.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|