Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest...

OLYMPUS Corporation of the Americas / initiated 2024-09-24 / Open, Classified / root cause: design

Olympus Corporation of the Americas began a recall of Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in... on . FDA classified it as Class II and gives the reason as "Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0232-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries.
Product code
OWN Confocal Optical Imaging
Reason for recall
Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
Root cause tag
design
FDA root cause
Component design/selection
Action
Olympus issued Urgent Medical Correction letter to Operating Room Director, Risk Management on 9/24/24. Letter states reason for recall, health risk and action to take: 1. Carefully read the content of this notification. 2. Quarantine and mark you Spectrum 3 unit appropriately to prevent usage. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification. 4. Acknowledge receipt of this letter through the Olympus recall portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number "0456 c. Complete the form as instructed. 5. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you report any complaints or any associated injuries to the Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at [email protected].
Quantity in commerce
3 units
Distribution
OH
Product codes and lots
UDI-DI: 08719324487047 Lot Code: Model: ART-SYS-0029, Serial: ART-SYS-2022-F04, ART-SYS-2022-A06, ART-SYS-2022-F05; Model: ART-SYS-0030; No distribution in the USA

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0232-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0232-2025Class IIOngoingVoluntary: Firm initiated