Z-0224-2025Class II

Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and...

Spectrum Medical Inc / initiated 2023-03-13 / Completed / root cause: design

Spectrum Medical Ltd. began a recall of Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a)... on . FDA classified it as Class II and gives the reason as "The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock". As of , the recall status on record is Completed, and MDM tags the root cause as design.

Recall record

Recall number
Z-0224-2025
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Spectrum Medical Inc
Product
Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.
Root cause tag
design
FDA root cause
Component design/selection
Action
All potentially affected product was identified. The firm corrected all products between 03/13/2024 and 08/07/2024 during service visits.
Quantity in commerce
172 units
Distribution
US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia.
Product codes and lots
UDI/DI 05060434420770: Serial Numbers: a) AU6004040, AU6004041, AU6004042, AU6004043, AU6004044, AU6004046, AU6004050, AU6004051, AU6004052, AU6004053, AU6004054, AU6004056, AU6004057, AU6004058, AU6004060, AU6004061, AU6004062, AU6004063, AU6004064, AU6004065, AU6004066, AU6004069, AU6004070, AU6004071, AU6004072, AU6004073, AU6004074, AU6004075, AU6004136, AU6004137, AU6004138, AU6004139, AU6004140, AU6004141, AU6004142, AU6004143, AU6004144, AU6004145, AU6004146, AU6004147, AU6004148, AU6004159, AU6004160, AU6004161, AU6004162, AU6004167, AU6004168, AU6004169, AU6004170, AU6004171, AU6004172, AU6004173, AU6004174, AU6004175, AU6004176, AU6004177, AU6004178, AU6004179, AU6004180, AU6004181, AU6004182, AU6004183, AU6004184, AU6004185, AU6004186, AU6004187, AU6004188, AU6004195, AU6004197, AU6004198, AU6004200, AU6004201, AU6004202, AU6004203, AU6004204, AU6004205, AU6004207, AU6004332, AU6004333, AU6004348, AU6004349, AU6004350, AU6004352, AU6004353, AU6004354, AU6004532, AU6004577, AU6004578, AU6004579, AU6004580, AU6004587, AU6004588, AU6004590, AU6004591, AU6004603, AU6004604, AU6004605, AU6004606, AU6004607, AU6004609, AU6004610, AU6004613, AU6004731, AU6004771, AU6004773, AU6004776, AU6004777, AU6004778, AU6004779, AU6004781, AU6004782, AU6004919, AU6004920, AU6004921, AU6004922, AU6004923, AU6004924, AU6004925, AU6004926, AU6004927, AU6004929, AU6004930, AU6004955, AU6004957, AU6004958, AU6004960, AU6004963, AU6004966, AU6004968, AU6004970, AU6004971, AU6004972, AU6005010, AU6005021, AU6005022, AU6005023, AU6005024, AU6005025, AU6005026, AU6005027, AU6005028, AU6005029, AU6005030, AU6005037, AU6005038, AU6005039, AU6005040, AU6005041, AU6005042, AU6005057, AU6005065, AU6005109, AU6005110, AU6005118, AU6005124, AU6005125, AU6005355, AU6006351; Serial Numbers: b) AU8001273, AU8001275, AU8001346, AU8001347, AU8001348, AU8001359, AU8001360, AU8001366, AU8001403, AU8001406, AU8001407, AU8001446, AU8001447, AU8001449

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0224-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0224-2025Class IICompletedVoluntary: Firm initiated