Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE...
Bard Peripheral Vascular Inc began a recall of Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF: MQ1210; BARD MARQUEE Kit 12G X... on . FDA classified it as Class II and gives the reason as "Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0467-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF: MQ1210; BARD MARQUEE Kit 12G X 13 CM, REF: MQK1213; BARD MARQUEE 12G X 13 CM, REF: MQ1213
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and misplacement, which could lead to tissue damage and bleeding.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- On 9/30/24, removal notices were sent to customers who were asked to do the following: 1) Discontinue use of product listed in the letter, and immediately dispose of all affected product remaining in your possession in accordance with your local facility s process. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, please share this notice to ensure that these entities are contacted and informed of the recall. 3) Complete and return the response form via email to [email protected] - If you purchased this product from a distributor, contact your distributor for further instructions and resolution. - Affected lots distributed from 05 April 2023 through 23 Aug 2024. If further assistance is required contact the firm: - North American Regional Complaint Center: 1-844-823-5433, Mon - Fri 8:00am and 5:00pm CT, [email protected] - Recall Related Questions: [email protected] On 1/15/2025, additional removal notices were mailed or emailed to distributors informing them of additional affected devices.
- Quantity in commerce
- 71,498
- Distribution
- Worldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC, OK, KY, AZ, IL, TX, NJ, NY, FL, TN, IN, OH, NC, CO, VA, GA, MD, CA, AL, WI, MO, IA, CT, NE, MI, MS, NV, KS, OR, SD, ID, NH, RI, WY, AR, MA, DE, ND, NM, AK, PR and the countries of Portugal, Spain, South Africa, Greece, United Arab Emirates, United Kingdom, Sweden, Finland, Belgium, Korea, Australia, Canada.
- Product codes and lots
- REF/UDI-DI/Lot(Expiration): MQK1210/00801741097263/1490758(9/28/2025), 1506044(1/28/2026), 1506045(1/28/2026), 1511827(2/28/2026), 1514692(3/28/2026), 1518977(4/28/2026), 1518978(4/28/2026), 1518979(4/28/2026), 1519143(4/28/2026), 1520507(5/28/2026), 1520508(5/28/2026), 1522157(5/28/2026), 1522158(5/28/2026), 1522577(5/28/2026), 1523785(6/28/2026), 1523914(6/28/2026), 1523915(6/28/2026), 1525370(6/28/2026), 1525371(6/28/2026), 1525373(6/28/2026), 1528417(7/28/2026), 1528418(7/28/2026), 1528495(7/28/2026), 1528496(7/28/2026), 1528497(7/28/2026), 1528560(7/28/2026), 0001618867(3/28/2028), 0001620281(3/28/2028), 0001632675(6/28/2028); MQ1210/00801741097119/1514460(3/28/2026), 1514461(3/28/2026), 1519144(4/28/2026), 1519145(4/28/2026), 1519456(4/28/2026), 1520066(5/28/2026), 1520067(5/28/2026), 1520506(5/28/2026), 1523783(6/28/2026), 1523784(6/28/2026), 0001618258(3/28/2028), 0001631611(6/28/2028), 0001633216(7/28/2028); MQK1213/00801741097270/1543517(10/28/2026), 1559312(2/28/2027), 1562724(3/28/2027), 1569737(4/28/2027), 1581652(7/28/2027); MQ1213/00801741097126/1558191(2/28/2027), 1559313(2/28/2027)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0467-2025 | Class II | Ongoing | Voluntary: Firm initiated |