Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0281-2025Class II

smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated...

Smith & Nephew, Inc. / initiated 2024-10-02 / Open, Classified / root cause: packaging

Smith & Nephew Inc began a recall of smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated knee femur prosthesis, metallic on . FDA classified it as Class II and gives the reason as "Complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0281-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Smith & Nephew, Inc.
Product
smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated knee femur prosthesis, metallic
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Smith & Nephew issued an "Urgent Medical Device RECALL Notice" to its consignees on 10/02/2024 via email. The notice explained the issue, potential risk of use of the affected product, and requested the return of the affected devices. Return affected product to: Smith+Nephew | Attn: Quality Hold and Field Action Returns Department | 1450 E Brooks Road, Building G | Memphis, TN 38116 Distributors, sales reps, and district offices were directed to notify their customers. If you have any questions or concerns regarding this recall, please contact FieldActions@smith nephew.com.
Quantity in commerce
5 units
Distribution
US Distribution to states of: MS, TN, NC, CA, GA, DE, AL, and OUS distribution to country of: Canada
Product codes and lots
UDI/DI 03596010469212, Batch number 23LM02331

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0281-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0281-2025Class IIOngoingVoluntary: Firm initiated
Z-0282-2025Class IIOngoingVoluntary: Firm initiated