smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated...
Smith & Nephew Inc began a recall of smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated knee femur prosthesis, metallic on . FDA classified it as Class II and gives the reason as "Complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0281-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Smith & Nephew, Inc.
- Product
- smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated knee femur prosthesis, metallic
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Smith & Nephew issued an "Urgent Medical Device RECALL Notice" to its consignees on 10/02/2024 via email. The notice explained the issue, potential risk of use of the affected product, and requested the return of the affected devices. Return affected product to: Smith+Nephew | Attn: Quality Hold and Field Action Returns Department | 1450 E Brooks Road, Building G | Memphis, TN 38116 Distributors, sales reps, and district offices were directed to notify their customers. If you have any questions or concerns regarding this recall, please contact FieldActions@smith nephew.com.
- Quantity in commerce
- 5 units
- Distribution
- US Distribution to states of: MS, TN, NC, CA, GA, DE, AL, and OUS distribution to country of: Canada
- Product codes and lots
- UDI/DI 03596010469212, Batch number 23LM02331
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0281-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0282-2025 | Class II | Ongoing | Voluntary: Firm initiated |