GE Healthcare / initiated 2024-10-03 / Open, Classified / root cause: software

GE Healthcare began a recall of GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal... on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0318-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, 105468-1-Centricity Universal Viewer Zero Footprint / ZFP-02254840, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096090, JP-340111-Centricity Universal Viewer Zero Footprint / ZFP-01275379, JP-275353-Centricity Universal Viewer Zero Footprint / ZFP-02042636; 6) 2089507-112, 3922-25-Centricity Universal Viewer Zero Footprint / ZFP-01290685, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, 1503586-Centricity Universal Viewer Zero Footprint / ZFP-21016603, RU0994-Centricity Universal Viewer Zero Footprint / ZFP-02331255; 7) 2089507-114, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-02360229, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-01803680, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225897, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225898, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225904, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01371500, M51725-Centricity Universal Viewer Zero Footprint / ZFP-01300873, M1781854-Centricity Universal Viewer Zero Footprint / ZFP-01553772, 609334-Centricity Universal Viewer Zero Footprint / ZFP-01282257, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, CMC-Centricity Universal Viewer Zero Footprint / ZFP-01820241, JP-100014-Centricity Universal Viewer Zero Footprint / ZFP-01890709, KW1102-Centricity Universal Viewer Zero Footprint / ZFP-02191178, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02327073, QA1114-Centricity Universal Viewer Zero Footprint / ZFP-02306397, SAU1433-Centricity Universal Viewer Zero Footprint / ZFP-01457278, TTSH-CVIS-Centricity Universal Viewer Zero Footprint / ZFP-02041523, NE4651-Centricity Universal Viewer Zero Footprint / ZFP-00038599, NE6703-Centricity Universal Viewer Zero Footprint / ZFP-00038552, NE6731-Centricity Universal Viewer Zero Footprint / ZFP-00038461, NE3006-Centricity Universal Viewer Zero Footprint / ZFP-00038362, NE2196-Centricity Universal Viewer Zero Footprint / ZFP-00038588, NPV738-Centricity Universal Viewer Zero Footprint / ZFP-00038637, NE2031-Centricity Universal Viewer Zero Footprint / ZFP-00038578, NE7405-Centricity Universal Viewer Zero Footprint / ZFP-00038372, NE2162-Centricity Universal Viewer Zero Footprint / ZFP-00038563, NE4878-
Product code
LLZ System, Image Processing, Radiological
Reason for recall
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to it consignees on 10/03/2024 via letter issued through a traceable means. The notice explained the issue and provided the following actions be to be implemented until the devices can be corrected: Option 1: Restrict access to these systems from the internet. or Option 2: Implement additional safeguards such as a VPN. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or a local Service Representative. GE HealthCare will correct all affected products at no cost. A GE HealthCare representative will make contact to arrange for the correction.
Quantity in commerce
120 units
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Argentina, Brazil, China, Dominican Republic, France, Germany, India, Indonesia, Ireland, Israel, Japan, Kuwait, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, Singapore, Spain, United Kingdom.
Product codes and lots
UDI/DI 00840682102988 for all models: 1) 2089507-083, Installed Product ID Numbers: GDZH-JINWAN-Centricity Universal Viewer Zero Footprint / ZFP-02171477, 103378-Centricity Universal Viewer Zero Footprint / ZFP-20039058, 1664823-Centricity Universal Viewer Zero Footprint / ZFP-20080654, 493356-Centricity Universal Viewer Zero Footprint / ZFP-20085691, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291850, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291877, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02292580, 81077-Centricity Universal Viewer Zero Footprint / ZFP-20680055, IL1469-Centricity Universal Viewer Zero Footprint / ZFP-02156638, JP-110145-Centricity Universal Viewer Zero Footprint / ZFP-01275393, 1333629-Centricity Universal Viewer Zero Footprint / ZFP-01300332; 2) 2089507-093, Installed Product ID Numbers: 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, Installed Product ID Numbers: JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, Installed Product ID Numbers: 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, Installed Product ID Numbers: 6) 2089507-112, Installed Product ID Numbers: 7) 2089507-114, Installed Product ID Numbers: 8) 2089507-115, Installed Product ID Numbers: 9) 2089507-116, Installed Product ID Numbers: 10) 2089507-117, Installed Product ID Numbers: 11) 2089507-118, Installed Product ID Numbers: 12) 2089507-119, Installed Product ID Numbers:

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0318-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0318-2025Class IIOngoingVoluntary: Firm initiated