IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
Siemens Healthcare Diagnostics, Inc. began a recall of IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282) on . FDA classified it as Class II and gives the reason as "The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0464-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
- Product code
- MTF Total, Prostate Specific Antigen (Noncomplexed & Complexed) For Detection Of Prostate Cancer
- Reason for recall
- The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- An Urgent Medical Device Correction letter dated 9/30/24 was sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, that may have been affected. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution IMMULITE 2000 and IMMULITE 2000 XPi kit lots D168 and above meet the IFU high-dose hook effect claim. Siemens continues to investigate root cause of the issue observed with the impacted kit lots. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please send a scanned copy of the completed form via email to [email protected]. Or to fax this completed form to the Customer Care Center at (312) 275-7795.
- Quantity in commerce
- 7,332 units
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- SMN 10706281: UDI - 00630414982212, Lot Number D164; SMN 10706282: UDI - 00630414982229, Lot Number D166; SMN 10380896 (OUS): UDI - 00630414961958; SMN 10380996 (OUS): UDI - 00630414961965; OUS Lot Numbers: 438, 439, 440
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0464-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.