Z-0568-2025Class II
Baxter MiniCap Extended Life PD Transfer Set (long) with Titanium Spike, Part Number T5C4325K; use in...
Baxter Healthcare Corporation / initiated 2024-10-23 / Open, Classified
Baxter Healthcare Corporation began a recall of Baxter MiniCap Extended Life PD Transfer Set (long) with Titanium Spike, Part Number T5C4325K; use in Peritoneal Dialysis on . FDA classified it as Class II and gives the reason as "There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0568-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter MiniCap Extended Life PD Transfer Set (long) with Titanium Spike, Part Number T5C4325K; use in Peritoneal Dialysis
- Product code
- KDJ Set, Administration, For Peritoneal Dialysis, Disposable
- Reason for recall
- There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.
- FDA root cause
- Under Investigation by firm
- Action
- Baxter issued a SAFETY ALERT notice to its consignees on 10/23/2024 via USPS first class mail. The notice explains the issue and potential hazard involved. The notice addressed to dialysis providers indicated that if a patient contacts your clinic experiencing separation of the transfer set, to replace their transfer set and contact Baxter. Notices addressed to peritoneal patients provided the following instructions: "If you experience a separation of the transfer set, please close the transfer set twist clamp and do not cut the patient line. Contact your doctor and/or nurse immediately to arrange for a transfer set replacement and further guidance." Distributors were directed to notify their customers. For general questions regarding this communication, please contact Baxter Renal Home Care Services which can be reached at 800-284-4060, option 1, between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday.
- Quantity in commerce
- 1536 units
- Distribution
- Worldwide distribution.
- Product codes and lots
- UDI/DI 0085412150710, Lot/Serial Numbers: H24E15066
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0559-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0560-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0561-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0562-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0563-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0564-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0565-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0566-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0567-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0568-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0569-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.