Z-0305-2025Class I

Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable...

Draeger, Inc. / initiated 2024-10-15 / Open, Classified / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable on . FDA classified it as Class I and gives the reason as "The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0305-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Draeger, Inc.
Product
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On October 15, 2024 Urgent Medical Device Correction letters were sent to customers. Necessary Action: The corrective action is to replace the ventilator motor assembly. You will be contacted by your local Draeger Service Representative to arrange a date for the ventilator motor assembly replacement. Until the corrective action is implemented, you may continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Monitor the patient s condition specifically their oxygenation status as a brief cessation of ventilation can potentially lead to e.g.; hypoxia, loss of lung recruitment, bradycardia and cardiac arrest. If preferred, the alarm priority of the "Ventilator error!!!" can be downgraded with "ALARM RESET" after switching to Man/Spont ventilation mode. Identification of affected medical devices: According to our records, you have received at least one Atlan anesthesia workstation manufactured by Draegerwerk AG & Co. KGaA that may be affected by this issue. Please refer to the attached list of affected serial numbers. Please ensure that all users of the Drger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. Please complete and return the attached Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them. If you have any questions regarding the operation of your Atlan A350/A350XL anesthesia workstation, please contact Draeger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 2 again). If you have any questions regarding
Quantity in commerce
455 units (310 units US, 145 OUS)
Distribution
US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
Product codes and lots
Model/Catalog Number: 8621600; UDI-DI Number: 04048675556183; Serial Numbers Distributed in the US: ASTA-0307; 0309; 0311; 0349; 0350; 0351; 0353; 0363; 0365; 0394; 0408; 0409 ASTB-0048; 0286; 0287; 0288; 0289; 0290; 0291; 0304; 0305; 0306; 0307; 0308; 0309; 0317; 0318; 0319; 0320; 0321; 0322 ASTC-0006; 0007; 0068; 0069; 0070; 0082; 0083; 0099; 0100; 0101; 0105; 0106; 0107; 0108; 0109; 0118; 0119; 0136; 0137; 0229; 0230; 0243; 0244; 0245 ASTD-0008; 0009; 0010; 0011; 0012; 0013; 0014; 0129; 0130; 0131; 0132; 0133; 0134; 0135; 0154; 0159; 0160; 0161; 0162; 0163; 0164; 0165; 0166; 0167; 0168; 0169; 0181; 0182; 0183; 0193; 0194; 0195; 0225; 0226; 0227; 0228; 0229; 0230; 0231; 0238; 0239; 0240; 0241; 0242; 0243; 0244; 0245 ASTE-0016; 0017; 0077; 0078; 0079; 0086; 0087; 0122; 0123; 0124; 0125; 0126; 0129; 0130; 0131; 0163; 0164; 0165; 0166; 0171; 0172; 0173; 0174; 0195; 0196; 0197; 0198; 0203; 0225; 0226; 0246; 0247; 0248; 0249; 0250; 0278; 0279; 0280; 0281; 0301; 0302; 0303; 0304; 0305; 0306; 0348; 0349; 0350; 0351; 0352; 0357; 0358; 0359; 0360; 0361; 0362 ASTF-0001; 0002; 0003; 0004; 0014; 0015; 0016; 0017; 0018; 0030; 0031; 0032; 0033; 0034; 0035; 0036; 0037; 0055; 0061; 0062; 0063; 0070; 0071; 0076; 0077; 0078; 0079; 0080; 0081; 0121; 0122; 0123; 0124; 0125; 0135; 0136; 0137; 0138; 0139; 0140; 0141; 0168; 0169; 0170; 0194; 0195; 0196; 0409; 0410; 0411; 0436; 0437; 0438; 0474; 0475; 0476; 0477; 0478 ASTH-0007; 0008; 0009; 0010; 0011; 0012; 0092; 0093; 0094; 0095; 0096; 0120; 0121; 0122; 0123; 0124; 0125; 0126; 0142; 0143; 0144; 0145; 0146; 0147; 0148; 0169; 0170; 0171; 0172; 0173; 0174; 0175; 0187; 0188; 0189; 0190; 0191; 0192; 0269; 0270; 0271; 0272; 0273; 0293; 0294; 0295; 0296; 0306; 0307; 0308; 0309; 0310; 0311; 0324; 0325; 0326; 0327; 0330; 0331; 0332; 0333; 0334; 0335; 0336; 0337; 0338; 0339; 0358; 0359; 0360; 0361; 0362; 0363; 0364; 0365; 0383; 0384; 0385; 0386; 0387; 0388; 0389; 0390; 0450; 0476; 0477; 0478; 0479; 0480; 0481; 0488; 0489; 0490; 0491 Serial Numbers Distributed OUS: ASTA-0253; 0289; 0290; 0327; 0362; 0404; 0405; 0426; 0427; 0428; 0429; 0430 ASTB-0003; 0005; 0007; 0028; 0029; 0030; 0031; 0032; 0033; 0075; 0076; 0077; 0079; 0080; 0081; 0082; 0083; 0084; 0134; 0195; 0206; 0219; 0220; 0237 ASTC-0001; 0002; 0003; 0004; 0026; 0138; 0163; 0180; 0181; 0194; ASTD-0015; 0016; 0050; 0061; 0062; 0063; 0064; 0065; 0066; 0203; 0204; 0205; 0265; 0266; 0267 ASTE-0034; 0035; 0057; 0058; 0059; 0060; 0062; 0085; 0108; 0127; 0128; 0175; 0202; 0223; 0224; 0300; 0325; 0326; 0327; 0328; 0329; 0353; 0354; 0355; 0356 ASTF-0051; 0052; 0053; 0054; 0112; 0113; 0435; 0471 ASTH-0039; 0040; 0045; 0046; 0081; 0082; 0176; 0193; 0194; 0204; 0217; 0218; 0219; 0225; 0226; 0243; 0244; 0245; 0246; 0255; 0256; 0382; 0403; 0404; 0405; 0406; 0434; 0435; 0436; 0437; 0451; 0452; 0453; 0 ASTJ-0063; 0064; 0065; 0066; 0067; 0068; 0089; 0090; 0091; 0092; 0145; 0146; 0147; 0171; 0237 ASTK-0002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0305-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0304-2025Class IOngoingVoluntary: Firm initiated
Z-0305-2025Class IOngoingVoluntary: Firm initiated