Z-0351-2025Class II

Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version...

OLYMPUS Corporation of the Americas / initiated 2024-09-13 / Open, Classified / root cause: labeling and UDI

Olympus Corporation of the Americas began a recall of Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product Description: Outer sheaths can be... on . FDA classified it as Class II and gives the reason as "Olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0351-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product Description: Outer sheaths can be combined with rigid Telescopes. Component: N/A
Product code
GCP Endoscope, Ac-Powered And Accessories
Reason for recall
Olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On October 11, 2024 URGENT: MEDICAL DEVICE CORRECTION letters were sent to consignees. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Therefore, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Examine your inventory and identify the above-listed device. 3. You should not use the Cystoscope Outer sheath, 22.5 French with a GreenLight Laser for BPH therapy until further notice from Olympus. Add a copy of this letter to any on-hand copies of the Instructions for Use. 4. Olympus requests that you acknowledge receipt of this letter. Acknowledge receipt of this letter through the Olympus web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number "0460 c. Complete the form as instructed. 5. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you report any complaints, including the device breaking off into the patient, or any associated injuries to the Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1), and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 from Monday through Friday, 9 am to 5 pm EDT, or by e-mail at [email protected].
Quantity in commerce
738 units
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Product codes and lots
Model: WA2280A, UDI: 04042761051729, Lot#: All lots.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0351-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0351-2025Class IIOngoingVoluntary: Firm initiated