Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0498-2025Class II

Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190...

OLYMPUS Corporation of the Americas / initiated 2024-09-18 / Open, Classified / root cause: labeling and UDI

Olympus Corporation of the Americas began a recall of Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A... on . FDA classified it as Class II and gives the reason as "To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0498-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
Product code
FDS Gastroscope And Accessories, Flexible/Rigid
Reason for recall
To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.
Root cause tag
labeling and UDI
FDA root cause
Reprocessing Controls
Action
On October 18, 2024 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. The range of affected serial numbers is included at the top of this letter. Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Implement use of the enclosed Addendum stating that the channel drying time has been extended by 12 minutes to purge water from the channels. This is accomplished by drying each channel for 3 minutes, in the sequence described in the addendum. Ensure all reprocessing personnel are completely knowledgeable and thoroughly trained on the new water purge requirements. 3. Olympus requests that you acknowledge receipt of this letter. Acknowledge receipt of this letter through the Olympus web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number "0461 c. Complete the form as instructed and include your account ID number: [XXXXXXXX] 4. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you report any complaints, including infections or persistent microbial colonization associated with any Olympus endoscope to the Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1), and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 from Monday through Friday, 9 am to 5 pm EDT, or by e-mail at [email protected].
Quantity in commerce
1749 units
Distribution
Domestic: US Nationwide Distribution.
Product codes and lots
Model: GIF-1TH190, UDI-DI: 04953170343360, Serial #: Less than 2446103.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0498-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0498-2025Class IIOngoingVoluntary: Firm initiated