smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
Smiths Medical ASD, Inc. began a recall of smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP on . FDA classified it as Class II and gives the reason as "Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0528-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
- Product code
- BSR Stylet, Tracheal Tube
- Reason for recall
- Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Smiths Medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 10/28/2024, mailed using a traceable means. The notice explained the issue, potential risk, and requested the consignee locate and destroy all affected products, and distribute the notice to all affected parties. Distributors were directed to notify their customers. For questions: About the communication: [email protected], 1-800-913-3897; To report adverse events: [email protected], 1-866-216-8806
- Quantity in commerce
- 10,223 units
- Distribution
- Worldwide distribution.
- Product codes and lots
- REF 14-504-68, UDI/DI 35021239002586, Lot Numbers: 0003393, 0003394, 0003357, 0003358, 0003267, 0003368, 0003634, 0003254, 0003369, 0003165, 0003184, 0003196, 0003204, 0003206, 0003226, 0003256, 0003300, 0003307, 0003308, 0003314, 0003322, 0003359, 0003360, 0003361, 0003362, 0003363, 0003364, 0003377, 0003380, 0003635, 0003643; REF 14-504-68JP, UDI/DI 35021239002586, Lot Numbers: 0003369, 0003377, 0003393, 0003394, 0003634, 0003635, 0003643
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0527-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0528-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0529-2025 | Class II | Ongoing | Voluntary: Firm initiated |