Z-0484-2025Class II

Acrobat-i Positioner. Model Number C-XP-5000Z

Maquet Cardiovascular LLC / initiated 2024-10-16 / Open, Classified / root cause: packaging

Maquet Cardiovascular, LLC began a recall of Acrobat-i Positioner. Model Number C-XP-5000Z on . FDA classified it as Class II and gives the reason as "Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0484-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Maquet Cardiovascular LLC
Product
Acrobat-i Positioner. Model Number C-XP-5000Z
Product code
MWS Stabilizer, Heart
Reason for recall
Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
Root cause tag
packaging
FDA root cause
Process change control
Action
MAQUET Cardiovascular LLC / Getinge notified consignees via letter FedEx 2-Day Delivery on 10/16/2024. Customers were instructed to forward the notification to all users and notify customers if further distributed, examine inventory and return any affected inventory, and to complete and return the MEDICAL DEVICE REMOVAL - RESPONSE FORM.
Quantity in commerce
14,338 units
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
Product codes and lots
Model Number C-XP-5000Z. UDI-DI Number: 00607567500006. Batch Numbers: 3000272452, 3000289270, 3000293675, 3000340401, 3000382289, 3000282355, 3000287205, 3000290368, 3000306921, 3000308238, 3000307458, 3000318694, 3000282741, 3000290350, 3000306394, 3000334128, 3000355318, 3000362345, 3000278266, 3000343467, 3000353846, 3000346421, 3000360292, 3000364706, 3000317546, 3000272461, 3000343116, 3000360175, 3000360569, 3000355540, 3000380141, 3000378810, 3000334322, 3000363049, 3000311731, 3000306920, 3000365139, 3000273855, 3000300092, 3000276339, 3000347351, 3000347358, 3000352249, 3000397151, 3000368288, 3000293958, 3000288356, 3000286557, 3000402809, 3000392590, 3000340972.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0484-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0484-2025Class IIOngoingVoluntary: Firm initiated
Z-0485-2025Class IIOngoingVoluntary: Firm initiated
Z-0486-2025Class IIOngoingVoluntary: Firm initiated