Z-0665-2025Class II
MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
Micro-X Ltd. / initiated 2023-11-28 / Open, Classified / root cause: design
Micro-X Ltd. began a recall of MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71 on . FDA classified it as Class II and gives the reason as "Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0665-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Micro-X Ltd.
- Product
- MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 11/28/23 Modification 17 was sent to customers telling them that Control Boards need to be replaced with Control Board PCBA, 12425-04 that will eliminate the issue. For technical assistance contact: [email protected], 1-206-249-8764 Firm asked distributors to do the following: On next visit please confirm that the bulletins were completed and return the documentation.
- Quantity in commerce
- 18
- Distribution
- US Nationwide Distribution: NJ, CA, PR
- Product codes and lots
- MXU-RV35 UDI-DI: 9357123000037 and MXU-RV71 UDI-DI: 935712300005 using Control Board PCBA, 12425-03 (or earlier). Serial Numbers: MSN00334, MSN00348, MSN00349, MSN00351, MSN00352, MSN00361, MSN00364, MSN00365,MSN00382, MSN00383, MSN00384, MSN00385, MSN00386, MSN00387, MSN00388, MSN00390, MSN00391, MSN00395
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0665-2025 | Class II | Ongoing | Voluntary: Firm initiated |