Z-0580-2025Class II

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Micro-X Ltd. / initiated 2024-10-22 / Open, Classified / root cause: manufacturing process

Micro-X Ltd. began a recall of MICRO-X Rover Mobile X-ray System, # MXU-RV35 on . FDA classified it as Class II and gives the reason as "Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0580-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Micro-X Ltd.
Product
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Product code
IZL System, X-Ray, Mobile
Reason for recall
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
Root cause tag
manufacturing process
FDA root cause
Environmental control
Action
On 10/22/24, Service Bulletin SB007-23-1.0 was emailed to customers restating receiving, storage, and operation conditions and informing customers that extreme humidity environments can adversely affect the equipment. For further information contact your local service representative: phone: 1-206-249-8764 Email: [email protected]
Quantity in commerce
6
Distribution
US: TX, PR, MD, CA
Product codes and lots
UDI-DI: 9357123000037, Serial Numbers: 348, 349, 361, 351, 365, 391

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0580-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0580-2025Class IIOngoingVoluntary: Firm initiated