Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile...
Securitas Healthcare LLC began a recall of Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application on . FDA classified it as Class II and gives the reason as "Certain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0570-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Securitas Healthcare LLC
- Product
- Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application
- Product code
- IQA System, Environmental Control, Powered
- Reason for recall
- Certain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- An URGENT: MEDICAL DEVICE CORRECTION letter dated 10/31/24 was sent to customers. Necessary Actions to be Taken by Clients Instructions for determining which versions of the Arial Server Software you are using are attached hereto as Exhibit C. Clients using Arial Server Software versions 11.1.4.9, 11.1.5.5, or 11.1.6.2 should: 1. Not decline previously cleared calls through the Arial App; 2. Restart the Arial Server Software if an end user has declined a previously completed call using the Arial App. After restarting the Arial Server Software, calls from the affected call point(s) will begin to be received by the Arial Server Software and the Arial Server Software will notify end users as intended. Instructions for restarting the Arial Server Software are included on Exhibit B. 3. Contact Securitas Healthcare Technical Support at 1 (800) 380-8883 to upgrade the Arial Server Software to version 11.1.7 or later. For general questions concerning this communication please contact Securitas Healthcare Technical Support at 1 (800) 380-8883. We ask that you please complete and return a copy of the acknowledgement form attached hereto as Exhibit A.
- Quantity in commerce
- 111 systems
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Switzerland.
- Product codes and lots
- All Systems using Arial Server Software versions 11.1.4 through 11.1.6 in conjunction with the Arial Mobile Application or Arial Care Giver Application
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|