Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR...
GE Medical Systems, LLC began a recall of Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR on . FDA classified it as Class II and gives the reason as "Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0520-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR,
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to [email protected]. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.
- Product codes and lots
- 1) UDI/DI To be provided, System ID 82427070262; 2) UDI/DI To be provided, System ID EX0133
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0517-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0518-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0519-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0520-2025 | Class II | Ongoing | Voluntary: Firm initiated |