Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0701-2025Class II

Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Multiport, Models: 1)...

Augustine Temperature Management, LLC / initiated 2024-12-04 / Open, Classified / root cause: design

Augustine Temperature Management, LLC began a recall of Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Multiport, Models: 1) Model # WC77, 2) Model # WC77D (Demo), 3) Model #... on . FDA classified it as Class II and gives the reason as "Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST"". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0701-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Augustine Temperature Management, LLC
Product
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Multiport, Models: 1) Model # WC77, 2) Model # WC77D (Demo), 3) Model # WC77R (Refurbished); used in conjunction with Warming Blankets and Mattresses.
Product code
DWJ System, Thermal Regulating
Reason for recall
Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".
Root cause tag
design
FDA root cause
Device Design
Action
Augustine Temperature Management issued a Field Notification to its consignees on 12/04/2024 via email. The notice explained the issue, potential risk, and advised that there is an update to the User and Technical Manual to further clarify existing instructions for the device. If the device was further distributed, those distributing the device were directed to notify those to whom the device was distributed. Consignees are expected to respond via email. An optional power board update is available. For questions or concerns, contact: [email protected] or by telephone, 952-465-3500, fax, 952-465-3501, 8AM - 5PM Central time.
Quantity in commerce
1776 units
Distribution
US and Canada
Product codes and lots
1) UDI-DI 00850011479025, Serial Numbers: 20080060-24060065; 2) UDI/DI 00850011479292, Serial Numbers: 20080060-24060065; 3) UDI-DI 00850011479308, Serial Numbers: 20080060-24060065

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0701-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0700-2025Class IIOngoingVoluntary: Firm initiated
Z-0701-2025Class IIOngoingVoluntary: Firm initiated
Z-0702-2025Class IIOngoingVoluntary: Firm initiated