Z-0667-2025Class I

Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia...

Draeger, Inc. / initiated 2024-10-15 / Open, Classified / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults... on . FDA classified it as Class I and gives the reason as "The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0667-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Draeger, Inc.
Product
Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On October 15, 2024 Urgent Medical Device Correction letters were sent to customers. Necessary Action: The corrective action is to replace the ventilator motor assembly. You will be contacted by your local Draeger Service Representative to arrange a date for the ventilator motor assembly replacement. Until the corrective action is implemented, you may continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Monitor the patient s condition specifically their oxygenation status as a brief cessation of ventilation can potentially lead to e.g.; hypoxia, loss of lung recruitment, bradycardia and cardiac arrest. If preferred, the alarm priority of the "Ventilator error!!!" can be downgraded with "ALARM RESET" after switching to Man/Spont ventilation mode. Please ensure that all users of the Draeger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. Complete and return the Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them. If you have any questions regarding the operation of your Atlan A350/A350XL anesthesia workstation, please contact Draeger Service Technical Support 8AM-8PM EST at 1-800-437-2437 (press 2, then 2, then 2 again). Questions regarding this Urgent Medical Device Correction notice, contact Michael Kelhart 8AM-4:30PM EST at 267-664-1131 or via email at [email protected].
Quantity in commerce
280 units (OUS only)
Distribution
No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
Product codes and lots
Model/Catalog Number: 8211300; UDI-DI Number: 04048675556343; Serial Numbers shipped directly from Draegerwerk AG & Co., KGaA to OUS: ASTA-0314; 0315; 0325; 0338; 0340; 0384; 0411; 0425 ASTB-0002; 0010; 0018; 0151; 0160; 0161; 0165; 0175; 0176; 0177; 0178; 0179; 0180; 0181; 0182; 0214; 0215; 0218; 0232; 0233; 0234; 0235; 0236; 0248; 0249; 0250; 0262; 0263; 0264; 0265; 0266; 0276; 0277; 0278; 0279; 0280 ASTC-0008; 0025; 0036; 0037; 0038; 0039; 0040; 0041; 0042; 0046; 0047; 0048; 0049; 0050; 0062; 0063; 0064; 0077; 0078; 0079; 0080; 0081; 0089; 0090; 0091; 0092; 0093; 0102; 0103; 0104; 0123; 0124; 0125; 0126; 0127; 0133; 0139; 0140; 0150; 0151; 0152; 0153; 0154; 0169; 0176; 0177; 0178; 0179; 0182; 0183; 0184; 0185; 0186; 0187; 0192; 0195; 0196; 0200; 0201; 0202; 0203; 0242 ASTD-0028; 0051; 0052; 0080; 0081; 0082; 0083; 0084; 0085; 0086; 0087 ASTE-0012; 0013; 0014; 0015; 0041; 0042; 0043; 0044; 0063; 0064; 0065; 0084; 0176; 0177; 0178; 0179; 0243; 0244; 0282; 0299 ASTF-0005; 0006; 0019; 0027; 0028; 0029; 0048; 0049; 0050; 0072; 0073; 0074; 0075; 0082; 0093; 0110; 0111; 0162; 0166; 0167; 0197; 0198; 0220; 0221; 0222; 0301; 0406; 0434 ASTH-0001; 0002; 0003; 0024; 0025; 0041; 0042; 0043; 0044; 0097; 0098; 0099; 0115; 0116; 0117; 0118; 0149; 0150; 0151; 0152; 0153; 0195; 0196; 0197; 0198; 0220; 0221; 0222; 0227; 0228; 0229; 0230; 0257; 0258; 0259; 0260; 0261; 0274; 0275; 0276; 0277; 0297; 0298; 0299; 0304; 0305; 0328; 0329; 0355; 0356; 0357; 0393; 0394; 0395; 0396; 0407; 0408; 0409; 0410; 0423; 0424; 0425; 0426; 0427; 0428; 0445; 0446; 0447; 0448; 0449; 0465; 0466; 0467; 0468; 0469; 0470; 0484; 0485; 0486; 0487 ASTK-0033; 0034; 0035; 0062

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0667-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0667-2025Class IOngoingVoluntary: Firm initiated
Z-0668-2025Class IOngoingVoluntary: Firm initiated