LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors...
Physio-Control, Inc. began a recall of LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO... on . FDA classified it as Class II and gives the reason as "Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1184-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Physio Control, Inc.
- Product
- LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 1/21/2025, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter via UPS 2-Day mail to customers to inform them that the LIFEPAK 15 V4 displays an error where "SpO2: Sensor does not support SpCO or SpMET". The error results in preventing users from utilizing LIFEPAK 15 V4 SpCO and SpMet functionalities when connected to specific lots of Masimo Rainbow Sensors. Customers are instructed to: 1. Immediately check your internal inventory to locate the product listed on the attached business reply form and remove them from their point of use. 2. Return the enclosed business reply form by email to [email protected] to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return and replacement of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please keep this communication close to the affected items until replacement. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. For questions or concerns - contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at [email protected].
- Quantity in commerce
- 622
- Distribution
- U.S. Nationwide distribution including in the states of CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MO, MS, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WV.
- Product codes and lots
- Catalog Number: 11996-000515 UDI-DI/GTIN code: 00843997010863 Lot Number: 23G79 Catalog Number: 11996-000519 UDI-DI/GTIN code: 00843997015608 Lot Numbers: 23HER 23HPT 23JBN 23JTA Catalog Number: 11996-000520 UDI-DI/GTIN code: 00843997015615 Lot Numbers: 23HNV 23JBP
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1184-2025 | Class II | Ongoing | Voluntary: Firm initiated |