Z-0583-2025Class II

Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344

Philips North America LLC / initiated 2024-11-16 / Open, Classified / root cause: software

Philips North America Llc began a recall of Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344 on . FDA classified it as Class II and gives the reason as "Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0583-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID software issue that may lead to misdiagnosis.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Philips notified consignees on about 11/16/2024, via certified mail. Consignees were instructed to identify any affected units on hand, identify if those units contain the affected software version, complete and return the response form, and circulate the Urgent Field Safety Notice Letter to all users of the device. To protect the user and patient from the issue of Plan Box not Updated, remain vigilant and confirm table position during procedures. Press any of the Emergency STOP buttons to stop unintended patient table motion and follow safety instructions in Section 6 of the IFU: Working with Specialized Exams. To avoid the issue of Wrong Patient ID with Anonymous patient, the operator should verify the patient information loaded into the patient data field is correct before scanning and rename the Anonymous record with the patient s name at the time of the scan.
Quantity in commerce
301 units
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, Israel, Italy, Japan, Jordan, Macao, Malaysia, Netherlands, New Zealand, Norway, Palestine, Panama, Poland, Portugal, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.
Product codes and lots
UDI-DI: (01)00884838101111. Software Version: 5.0.X. Model Number (Serial Numbers): 728333 (10266, 10261, 10264, 10219, 10265, 10197, 10182, 10175, 10192, 10179, 10196, 10180, 10188, 10190, 10181, 10183, 10178, 10186, 10185, 10199, 10184, 10198, 10193, 10187, 10189, 10195, 10191, 10194, 10018, 10023, 10027, 10062, 10072, 10078, 10079, 10080, 10084, 10088, 10101, 10104, 10112, 10117, 10113, 10116, 10032, 10049, 10133, 10140, 10152, 10153, 10166, 10170, 10171, 10173, 10172, 10174, 10016, 10015, 10017, 10019, 10022, 10021, 10020, 10025, 10024, 10033, 10031, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10043, 10042, 10014, 10044, 10046, 10048, 10045, 10047, 10051, 10050, 10053, 10055, 10052, 10056, 10058, 10057, 10059, 10060, 10061, 10065, 10066, 10068, 10067, 10070, 10071, 10063, 10069, 10074, 10075, 10077, 10076, 10081, 10082, 10087, 10089, 10085, 10086, 10091, 10094, 10093, 10095, 10096, 10090, 10099, 10098, 10102, 10100, 10103, 10097, 10105, 10106, 10107, 10108, 10109, 396001, 396003, 10110, 10111, 10114, 10115, 10122, 10118, 10120, 10121, 10123, 10119, 10124, 10125, 10126, 10127, 10129, 10128, 10131, 10130, 10132, 10135, 10136, 10134, 10137, 10139, 10138, 10142, 10141, 10143, 10144, 10145, 10146, 10147, 10149, 10150, 10151, 10155, 10156, 10159, 10148, 10160, 10161, 10162, 10158, 10164, 10163, 10165, 10167, 10168, 10169, 10013, 10177, 10176, 10210, 10218, 10217, 10206, 10202, 10200, 10211, 10207, 10204, 396004, 10208, 10201, 10205, 10203, 10212, 10213, 10225, 10214, 10220, 10215, 10222, 10227, 10247, 10209, 10226, 10230, 10228, 10232, 10231, 10229, 10221, 10233, 10234, 10235, 10237, 10238, 10223, 10239, 10241, 10242, 10236, 10245, 396005, 10248, 10249, 10243, 10244, 10250, 10259, 10258, 10157, 10262, 10246, 10255, 10254, 10240, 10251, 10256, 10224, 10253, 10257, 10260, 10263), 728340 (397060, 397061, 397022, 397023, 397020, 397024, 397021, 397025, 397027, 397031, 397026, 397033, 397032, 397030, 397028, 397029, 397015, 397016, 397004, 397007, 397005, 397002, 397006, 397008, 397010, 397003, 397011, 397013, 397009, 397012, 397014, 397017, 397019, 397018, 397036, 397037, 397038, 397034, 397041, 397040, 397039, 397044, 397046, 397043, 397045, 397047, 397048, 397049, 397053, 397050, 397054, 397055, 397056, 397058, 397059), 728344 (397057).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0583-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0583-2025Class IIOngoingVoluntary: Firm initiated