Z-0727-2025Class II

Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of...

Siemens Healthcare Diagnostics Inc. / initiated 2024-11-18 / Open, Classified / root cause: design

Beckman Coulter, Inc. began a recall of Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum... on . FDA classified it as Class II and gives the reason as "Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0727-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma.
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay carryover due to insufficient washing. The contamination can cause a falsely elevated Access hsTnI result in subsequent sample(s) which can impact patient care (potentially but not limited to unnecessary coronary imaging or diagnostic catheterization) if the result is near the medical decision points.
Root cause tag
design
FDA root cause
Component design/selection
Action
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 11/18/2024 via USPS and email. The notice explained the issue, impact to patient care, and requested the following actions be taken: " - Follow your established laboratory protocols for analyzing and retesting discrepant samples if an observed Access hsTnI test result does not align with the patient s clinical presentation. - Beckman Coulter recommends reviewing the content of this letter with your laboratory and/or medical director to determine appropriate next steps. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded the affected product listed above to another laboratory, please provide them a copy of this letter."
Quantity in commerce
2,109,804 units
Distribution
Worldwide
Product codes and lots
UDI/DI 15099590693183, All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0727-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0727-2025Class IIOngoingVoluntary: Firm initiated