Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0695-2025Class II

HUDSON RCI mBrace, ET Tube Holder and 2 Point Head Strap, REF DYNJMBRC2; tracheal tube fixation device

MEDLINE INDUSTRIES, LP - Northfield / initiated 2024-11-18 / Open, Classified / root cause: manufacturing process

MEDLINE INDUSTRIES, LP - Northfield began a recall of HUDSON RCI mBrace, ET Tube Holder and 2 Point Head Strap, REF DYNJMBRC2; tracheal tube fixation device on . FDA classified it as Class II and gives the reason as "Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0695-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
HUDSON RCI mBrace, ET Tube Holder and 2 Point Head Strap, REF DYNJMBRC2; tracheal tube fixation device
Product code
CBH Device, Fixation, Tracheal Tube
Reason for recall
Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medline Industries issued an URGENT MEDICAL DEVICE RECALL to its consignees of 11/18/2024 via USPS and email. The notice explained the issue, risk to user, and requested the destruction of the affected product. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Quantity in commerce
1332 units
Distribution
US and Canada
Product codes and lots
UDI/DI 20193489005308, (case), 10193489005301 (each); All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0695-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0695-2025Class IIOngoingVoluntary: Firm initiated
Z-0696-2025Class IIOngoingVoluntary: Firm initiated
Z-0697-2025Class IIOngoingVoluntary: Firm initiated
Z-0698-2025Class IIOngoingVoluntary: Firm initiated