Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216

Philips North America LLC / initiated 2024-11-20 / Open, Classified / root cause: software

Philips North America Llc began a recall of Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216 on . FDA classified it as Class II and gives the reason as "The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0686-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Consignees were notified on about 11/20/2024 via certified mail. Consignees were instructed to identify affected units, share the communication with clinical staff, and place the Urgent Medical Device Correction with the documentation of the Philips Smart-hopping 2.0 AP 1.4 GHz and associated devices. A Philips representative will contact customers to arrange a software upgrade to E.19.02 (5.0.19.02) or greater to correct the issues. Distributors were instructed to contact customers and provide the Medical Device Correction to each.
Quantity in commerce
4,811 units
Distribution
Nationwide distribution
Product codes and lots
GTIN: 00884838099876. Model Number: 867216. All software versions are affected. UDI Numbers: (01)00884838099876(21)DK34706344, (01)00884838099876(21)DK33603995, (01)00884838099876(21)DK34706512, (01)00884838099876(21)DK33402245, (01)00884838099876(21)DK34706012, (01)00884838099876(21)DK33402324, (01)00884838099876(21)DK33603538, (01)00884838099876(21)DK34405346, (01)00884838099876(21)DK32801033, (01)00884838099876(21)DK34705955, (01)00884838099876(21)DK40307155, (01)00884838099876(21)DK34405363, (01)00884838099876(21)DK33604046, (01)00884838099876(21)DK34706011, (01)00884838099876(21)DK34706010, (01)00884838099876(21)DK33402486, (01)00884838099876(21)DK34405820, (01)00884838099876(21)DK34405539, (01)00884838099876(21)DK34405788, (01)00884838099876(21)DK33403107, (01)00884838099876(21)DK33603456, (01)00884838099876(21)DK34706099, (01)00884838099876(21)DK32801250, (01)00884838099876(21)DK32801095, (01)00884838099876(21)DK33604249, (01)00884838099876(21)DK33403062, (01)00884838099876(21)DK34706432, (01)00884838099876(21)DK33403214, (01)00884838099876(21)DK33402554, (01)00884838099876(21)DK40307129, (01)00884838099876(21)DK33603678, (01)00884838099876(21)DK32801039, (01)00884838099876(21)DK33402660, (01)00884838099876(21)DK34706048, (01)00884838099876(21)DK33904683, (01)00884838099876(21)DK33402439, (01)00884838099876(21)DK34405702, (01)00884838099876(21)DK40306709, (01)00884838099876(21)DK32800996, (01)00884838099876(21)DK34405910, (01)00884838099876(21)DK32800867, (01)00884838099876(21)DK33603918, (01)00884838099876(21)DK33604048, (01)00884838099876(21)DK33904818, (01)00884838099876(21)DK34706539, (01)00884838099876(21)DK34405884, (01)00884838099876(21)DK33403219, (01)00884838099876(21)DK33603727, (01)00884838099876(21)DK34706371, (01)00884838099876(21)DK33402675, (01)00884838099876(21)DK33402648, (01)00884838099876(21)DK33604588, (01)00884838099876(21)DK33604264, (01)00884838099876(21)DK33402127, (01)00884838099876(21)DK34706378, (01)00884838099876(21)DK32801318, (01)00884838099876(21)DK33403095, (01)00884838099876(21)DK33603562, (01)00884838099876(21)DK34706181, (01)00884838099876(21)DK33604554, (01)00884838099876(21)DK33904888, (01)00884838099876(21)DK33402343, (01)00884838099876(21)DK40306696, (01)00884838099876(21)DK40306992, (01)00884838099876(21)DK33603488, (01)00884838099876(21)DK32800866, (01)00884838099876(21)DK34706255, (01)00884838099876(21)DK34706320, (01)00884838099876(21)DK34405767, (01)00884838099876(21)DK34706095, (01)00884838099876(21)DK33402405, (01)00884838099876(21)DK33603933, (01)00884838099876(21)DK33604442, (01)00884838099876(21)DK40306831, (01)00884838099876(21)DK33403181, (01)00884838099876(21)DK34405435, (01)00884838099876(21)DK34405878, (01)00884838099876(21)DK33905047, (01)00884838099876(21)DK33403263, (01)00884838099876(21)DK33402108, (01)00884838099876(21)DK33904987, (01)00884838099876(21)DK33402141, (01)00884838099876(21)DK33402149, (01)00884838099876(21)DK40306877, (01)00884838099876(21)DK33603419, (01)00884838099876(21)DK33402615, (01)00884838099876(21)DK33604419, (01)00884838099876(21)DK32801254, (01)00884838099876(21)DK32801299, (01)00884838099876(21)DK33604191, (01)00884838099876(21)DK33403177, (01)00884838099876(21)DK33904965, (01)00884838099876(21)DK34405697, (01)00884838099876(21)DK33402251, (01)00884838099876(21)DK33402899, (01)00884838099876(21)DK33604612, (01)00884838099876(21)DK33905057, (01)00884838099876(21)DK32801016, (01)00884838099876(21)DK33403251, (01)00884838099876(21)DK32801153, (01)00884838099876(21)DK40306893, (01)00884838099876(21)DK33403133, (01)00884838099876(21)DK33904707, (01)00884838099876(21)DK33402794, (01)00884838099876(21)DK34405477, (01)00884838099876(21)DK40306707 ,(01)00884838099876(21)DK32800977, (01)00884838099876(21)DK33603959, (01)00884838099876(21)DK33603543, (01)00884838099876(21)DK33402371, (01)00884838099876(21)DK34405486, (01)00884838099876(21)DK33403272, (01)00884838099876(21)DK33603742, (01)00884838099876(21)DK32801045, (01)00884838099876(21)DK33604374, (01)00884838099876(21)DK32801218, (01)00884838099876(21)DK33604349, (01)00884838099876(21)DK33905158, (01)00884838099876(21)DK33603443, (01)00884838099876(21)DK34706451, (01)00884838099876(21)DK33402451, (01)00884838099876(21)DK34405777, (01)00884838099876(21)DK34706063, (01)00884838099876(21)DK33402692, (01)00884838099876(21)DK34706448, (01)00884838099876(21)DK33604062, (01)00884838099876(21)DK32800961, (01)00884838099876(21)DK33604477, (01)00884838099876(21)DK33904802, (01)00884838099876(21)DK40307072, (01)00884838099876(21)DK32801156, (01)00884838099876(21)DK32801317, (01)00884838099876(21)DK32800916, (01)00884838099876(21)DK33603989, (01)00884838099876(21)DK33905142, (01)00884838099876(21)DK33904796, (01)00884838099876(21)DK33604177, (01)00884838099876(21)DK33603571, (01)00884838099876(21)DK33604137, (01)00884838099876(21)DK33604529, (01)00884838099876(21)DK33604398. .

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0686-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0686-2025Class IIOngoingVoluntary: Firm initiated