Z-0776-2025Class II
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product Code 35700BAX2
Baxter Healthcare Corporation / initiated 2024-11-19 / Open, Classified / root cause: manufacturing process
Baxter Healthcare Corporation began a recall of Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product Code 35700BAX2 on . FDA classified it as Class II and gives the reason as "Devices were identified as released after repair without full testing being performed, which includes flow testing". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0776-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product Code 35700BAX2
- Product code
- FRN Pump, Infusion
- Reason for recall
- Devices were identified as released after repair without full testing being performed, which includes flow testing.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Customers were contacted via telephone beginning 11/19/24. Baxter Healthcare is calling to notify you of a potential issue with the Spectrum infusion pump. One of your devices has been identified as released after repair without full testing being performed, which includes flow testing. Baxter is requesting the return of the impacted infusion pump to perform all tests required and confirm that the infusion pump performs as expected. As part of this process, we need to make you aware of the potential hazard associated with flow related issues. Flow issues may cause excessive therapy, free-flow condition, insufficient therapy or interruption of therapy. These may lead to serious adverse health consequences, including permanent injury and death. Send signed form to Global CORP Service Quality Field Actions [email protected]
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution in the states of AL, IN.
- Product codes and lots
- GTIN 00085412498683; Serial Number: 2153638
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0775-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0776-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.