Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and...
Augustine Temperature Management, LLC began a recall of Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses on . FDA classified it as Class II and gives the reason as "HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST"". As of , the recall status on record is Completed, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0841-2025
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Augustine Temperature Management, LLC
- Product
- Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses.
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.
- Quantity in commerce
- 77 units
- Distribution
- Worldwide
- Product codes and lots
- UDI-DI 00850011479582, Serial Numbers: 23010166, 23010167, 23010168, 23010169, 23010170, 23010171, 23010172, 23010173, 23010174, 23010175, 23010176, 23010177, 23010178, 23010179, 23010180, 23010181, 23010182, 23010183, 23010184, 23010185, 24020001, 24020002, 24020003, 24020004, 24020005, 24020006, 24030001, 24030002, 24030003, 24030004, 24030005, 24030021, 24030022, 24030023, 24030024, 24030025, 24030026, 24030027, 24030028, 24030029, 24030030, 24030016, 24030017, 24030018, 24030019, 24030020, 22120033, 24030011, 24030012, 24030013, 24030014, 24030015, 24030006, 24030007, 24030008, 24030009, 24030010, 24060001, 24060002, 24060003, 24060004, 24060005, 24060006, 24060007, 24060008, 24060009, 24060010, 24060024, 24060025, 24060026, 24060027, 24060028, 24060011, 24060012, 24060013, 24060014, 24060015.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0839-2025 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0840-2025 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0841-2025 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0842-2025 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-0843-2025 | Class II | Completed | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.