Z-0809-2025Class II

Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with...

GE Healthcare / initiated 2024-11-25 / Open, Classified / root cause: labeling and UDI

Spacelabs Healthcare, Inc. began a recall of Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data on . FDA classified it as Class II and gives the reason as "Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0809-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 11/25/2024, correction notices were sent to customers asking them to do the following: 1) Service manual has been updated by adding a section for "Required periodic system maintenance of Spacelabs Telemetry Farm", pages 5-30 through 5-35. 2) Download current versions of the Service Manual and PM Instructions at https://manuals.spacelabshealthcare.com/ 3) You may also obtain a copy of the latest manual by contacting the firm's Technical Support. 4) Discard previous copies of these documents 5) Complete the online customer acknowledgement form which can be accessed by using URL: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-windows-web-proxy-issue/ For additional information or to report any adverse events as a result of this issue, please contact the firm's Technical Support 1-800-522-7025 and select 2 for Technical Support
Quantity in commerce
2265
Distribution
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic
Product codes and lots
Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0809-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0809-2025Class IIOngoingVoluntary: Firm initiated