Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1172-2025Class II

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction...

Augustine Temperature Management, LLC / initiated 2024-02-15 / Open, Classified / root cause: design

Augustine Temperature Management, LLC began a recall of Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses on . FDA classified it as Class II and gives the reason as "Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1172-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Augustine Temperature Management, LLC
Product
Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.
Product code
DWJ System, Thermal Regulating
Reason for recall
Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
Root cause tag
design
FDA root cause
Device Design
Action
Augustine Temperature Management contacted their consignees by visit beginning on 02/15/2024 to replace the affected cables. The firm collected email confirmation that the original cables had either been returned or destroyed. If you have any questions, contact the Product Manager at 952-465-3526 or email:
Quantity in commerce
7 unit
Distribution
US distribution to states of: AL, TX, NV
Product codes and lots
UDI-DI 00850011479674, Lot Number 2339

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1172-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1171-2025Class IIOngoingVoluntary: Firm initiated
Z-1172-2025Class IIOngoingVoluntary: Firm initiated