Z-0812-2025Class II

GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2)...

GE Healthcare / initiated 2024-11-25 / Open, Classified / root cause: design

GE Medical Systems, LLC began a recall of GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0812-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided
Product code
IZL System, X-Ray, Mobile
Reason for recall
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Root cause tag
design
FDA root cause
Component design/selection
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/25/2024 via letter issued using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by the user: "1. Perform a check of your device prior to each patient exam to look for the following behaviors. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. . Column vertical movement requires increased effort. . Column has change in resistance (i.e. movement is not smooth) during vertical movement. . Column makes abnormal noises during vertical movement. 2. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. 3. Please retain this document for your records. 4. Please complete and return the attached acknowledgement form to [email protected]" For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. On 02/11/2025, the firm send an updated notice with a corrected GTIN for Optima XR200amx.
Quantity in commerce
281 units
Distribution
Worldwide
Product codes and lots
1) 5555000-12: UDI/DI none, Serial Numbers: 082407100532; UDI/DI to be provided, Serial Numbers: 082407160512; 2) 5555000-5: UDI/DI 840682115261, Serial Numbers: 1557XR0205, 352867XRA, 478765CDP1, 509962KV220, 601358OPTX220, 620421P3, 813872OPT3, SB2201700099WK; UDI/DI none, Serial Numbers: 479314220, 00028MOB12, 203694SB3, 205820DP1, 208666KDM, 209334OPT1, 210703M3, 240566PX2, 240964OPT1, 248465DP3, 256HHPORT2, 256HHPORT3, 256MHPORT1, 281MSL220, 281MSL2202, 305KEN220A, 306766DMOB2, 310680PORTDR2, 314251NAV7, 325574OPTITBD2, 330489DIGPORT, 333OPTI2, 333OPTI3, 333OPTI4, 352596OPT1, 352LEEDP1, 352VILDP1, 386FHODP3, 401OPTI1, 405636P2, 405636P240, 405752OPT1, 405752OPT2, 405752OPT4, 405DELCITY220, 407303OPTIMAX220, 408259AP6, 417347DPORT1, 417556P1, 417556P2, 417556P3, 419251DP1, 419251DP2, 419383DP1, 434654OPTIMA, 435CVXR220, 480MAYO220A, 501364OPTI1, 501364OPTI2, 509474DR1, 510437OPT1, 510727OPT3, 512SAMCSB, 513870DP1, 520TMCOPT2, 541426XR220, 559447OPT1, 559784OPT2, 561694OPT1, 561694OPT2, 561965DPORT5, 561965DPORT7, 573882DP1, 573882DP2, 573882DP3, 580421OPT1, 601288SB1, 601288SB2, 601288SB3, 601288SB6, 601928D220, 608364OPMX220, 613938XR3, 626457GEPORT1, 703369MDIG2, 703664M7, 706596SFP2, 706651DP, 715838DP2, 718206OPT240C, 718267DP5, 718670PTDR1, 720UCHSB7, 727848SB1, 731541SB1, 731541SB1A, 731541SB2, 731541SB3, 732324PA1, 732828DP11, 757388DP4, 770385220D1, 772468OPT1, 785776AMX220P1, 804281SMDM1, 806296OPTI220, 813302OPT3, 814837OPT, 815300M220C, 817927JPSXR220G, 817MCAXR220B, 828213DXP1, 828262P2202, 830RHHM1, 830RHHM2, 843847220B, 850325DP2, 863699OP220, 903886XR220, 905527XR220, 905845MOB4, 908685DP2, 909464220P1, 910615OPTIMA1, 910615OPTIMA2, 915521OPTIMA1, 915595OPT1, 940898XR220A, 940898XR240D, 985867M1, AXXXXX38, ORMCPORTA, ORMCPORTB, ORMCPORTC, ORMCPORTD, TAMXDRP03, WD0005, XR1036454WK5; 3) 5555000-6: UDI/DI 840682115261, Serial Numbers: 06051MOB06, 228809XR220, 228818XR220, 270798XR220B, 281KINGOPT3, 303450SB3, 303695SB3, 416323MOB3, 530888OPT1, 540981ODP3R, 540981ODP9, 615769OPTX220, 713529MDR220, 717782OPT4, 7982XR0192, 802775SB1, 813681OPT2, 918307OPT2, MHSWOPT2201, SA2513MM03, SB2201600071WK, SB2201700085WK, SJRO220P1, SJRO220P3, UTAHJAZZ, XR1864CO; UDI/DI none, Serial Numbers: 083007001584712, 0850070865, 0850070866, 0850070870, 0850070913, 0856070254, 0856070267, 0856070268, 091026DEMO2245, 10181MOB16, 140024RX14, 206598USB7, 209943OTP1, 214PH220G, 228575D220, 254HHSMOB1, 256429PORT2, 270688SB5, 281338BAPOR1, 281338CL221, 281338MDR220, 281348KWDM2, 281348KWDM3, 281KINGOPT1, 281MWB2202, 281MWB2203, 304255OPT220, 304623OPT220, 308865GSDP1, 309779RDIG2, 314525NAV15, 314525NAV16, 318675PW1, 318675PW2, 318769RDPT1, 318769RDPT2, 330480DP1, 330480DP2, 330480DP3, 330480DP4, 370OPTI1, 409938DR1220, 412232DMX2, 412372DMX1, 412647DMX2, 419291DP4, 419893DP1, 423493DP2, 432685OPT2, 434799GE220, 469764XR220H, 504454XR221, 504785MD1, 504873TMD8, 513732DP1, 563742NDIG1, 574267XR220, 586263DP1, 586493DP1, 601798D220, 614257DP4, 615284STMDR2, 615284STMDR3, 615316DP1, 615342DP1, 615391SUR, 702SIEXR221, 702SIEXR226, 713359BSDM3, 713440DP1, 713441OPCDX1, 714966GEPORT2, 720UCHSB9, 740592OPTIMA2, 77522MOB14, 801479SB2, 810342DP2, 817255NHXR220A, 817HUG220C, 817PLAZA220A, 817SAGERXR220A, 865541DP2, 865545DP1, 865545DP3, 903CSM222, 904OPMCDP3, 914737PORTWK3, 915577XR220, 916UCDAMXDR23, 916UCDAMXDR27, 918579OPT2, 919774DIGPRT, 928757XR222, 936539DM1, 936539DM2, 949517DR1, 949764DR2, 949764DR3, 949764DR4, 972420XR220B, 972420XR220C, 972566XR220F, 979MGXR220, LLPM220P1, MHKOPT220, SA1056MM06, SA2161MM04, XR1031657WK8, ZA1067RX30; 4) A0659PB: UDI/DI 840682115261, Serial Numbers: 702880EROP1; UDI/DI none, Serial Numbers: 206598USB6; 5) to be provided: UDI/DI none, Serial Numbers: 240826OPT1, 772SRDP1, 817HUG220B 860545SB2A; UDI/DI to be provided, Serial Numbers: 907852OPT22019, SA1075MM04.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0812-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0811-2025Class IIOngoingVoluntary: Firm initiated
Z-0812-2025Class IIOngoingVoluntary: Firm initiated
Z-0813-2025Class IIOngoingVoluntary: Firm initiated