EMPOWR 3D Knee Tibial Insert, 5R 16MM, VE, REF: 342-16-705; EMPOWR 3D Knee Tibial Insert, 3R 14MM, VE, REF...
Encore Medical, LP began a recall of EMPOWR 3D Knee Tibial Insert, 5R 16MM, VE, REF: 342-16-705; EMPOWR 3D Knee Tibial Insert, 3R 14MM, VE, REF: 342-14-703 on . FDA classified it as Class II and gives the reason as "5R 16MM knee tibial insert package may contain knee tibial insert 3R 14MM and vice versa, which may cause surgical delay to get a new product". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0766-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- EMPOWR 3D Knee Tibial Insert, 5R 16MM, VE, REF: 342-16-705; EMPOWR 3D Knee Tibial Insert, 3R 14MM, VE, REF: 342-14-703
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- 5R 16MM knee tibial insert package may contain knee tibial insert 3R 14MM and vice versa, which may cause surgical delay to get a new product.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On 12/18/2024, recall notices were emailed to consignees who were asked to do the following: 1. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 2. Review your stock for the products and lot numbers for the items to be returned. 3. Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service. 4. Complete and return the acknowledgement and response form via email to [email protected] If you have questions contact the firm at [email protected]
- Quantity in commerce
- 33
- Distribution
- US Nationwide distribution in the states of VA, MN, TX, WI, IA, IL, IN, SC, NY, MD, CA, FL, AZ, WA.
- Product codes and lots
- REF/UDI-DI/Lot: 342-16-705/00888912167628/159T1086 342-14-703/00888912167505/137T1086