Z-1020-2025Class II

HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic

Instrumentation Laboratory Co. / initiated 2024-12-18 / Open, Classified / root cause: labeling and UDI

Instrumentation Laboratory began a recall of HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic on . FDA classified it as Class II and gives the reason as "Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1020-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Instrumentation Laboratory Co.
Product
HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic
Product code
KFF Assay, Heparin
Reason for recall
Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On December 18, 2024 URGENT MEDICAL DEVICE CORRECTION letters were email to customers. Mandatory Customer Actions Based on the above, please take the following immediate actions: - Check inventory for lots of HemosIL Heparin Calibrators (Part No. 0020300600). See Appendix A for a list of in-date lots with corrected value assignments for the HEP CAL 2 and HEP CAL 3 levels. - Follow the instructions below to update the ACL TOP test parameters (same instructions for all models): - Select Setup from the menu bar and then select Test List from the dropdown menu. - From Test List , double click on the Test Code Anti-Xa to open Test Definition Anti-Xa . - From Test Definition Anti-Xa, click on Pre-diluted calibrators under the header Calibration . An editable table titled Calibration points will appear with HEP CAL 1, HEP CAL 2, and HEP CAL 3 and their predefined Target Values IU/mL of 0.00, 0.80 and 2.00, respectively. - In the Calibration points table, change the target values for HEP CAL 2 and HEP CAL 3 to the lot-specific calibrator values provided in Appendix A. - After entering the corrected HEP CAL 2 and HEP CAL 3 values, click on the save icon (picture of floppy disk) to save the changes. - Recalibrate using the HemosIL Liquid Anti-Xa assay lot in use after making the test parameter update. - Run quality controls per labeled instructions for use. If needed, determine control ranges for HemosIL UF Heparin Controls (Part No. 0020300300) and HemosIL Low Molecular Weight Heparin Controls (Part No. 0020300200) in accordance with your internal protocol until the new acceptance ranges are available. - Repeat the above test parameter update, recalibration, and quality control for all ACL TOP instruments. - Share this information with your laboratory staff. - Forward this notification to all affected locations within your facility. - Retain a copy of this notification for your records. - Complete and sign the enclosed Mandatory Response Tracking Form a
Quantity in commerce
10,952 units (5,303 US, 5,649 OUS) (Additional 11,562 units updated 4/15/2026)
Distribution
Worldwide - US Nationwide and the countries of Andorra, Algeria, Armenia, Australia, Austria, Belarus, Belgium, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Cyprus, Czechia, Denmark, Finland, France, Georgia, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Korea (the Republic of), Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Macao, Malaysia, Mexico, Netherlands (the), Norway, Philippines (the), Poland, Portugal, Qatar, Romania, Russian Federation (the), Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan (Province of China), Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Product codes and lots
Part Number: 0020300600; UDI-DI: 08426950472483; Lot Numbers: N0128933, N0220450, N0522479, N0724070, N1026509, N0239584, N0531891, N0733658, N0249207, N0442161, N0744673; ****Lot numbers added 4/15/2026**** N0531891, N0733658, N0249207, N0442161, N0744673, N1147734, N020463, N0250464, N0452277, N0553155, N0754506;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1020-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1020-2025Class IIOngoingVoluntary: Firm initiated