Z-1120-2025Class II

(1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) BD...

CareFusion 303, Inc / initiated 2025-01-08 / Open, Classified / root cause: labeling and UDI

CareFusion 303, Inc. began a recall of (1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) BD Pyxis MedStation ES Tower, Catalog No. 352; (4) BD... on . FDA classified it as Class II and gives the reason as "Labeling is insufficient regarding delays in access to medication. The firm is strengthening the label as an additional risk mitigation". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1120-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
(1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) BD Pyxis MedStation ES Tower, Catalog No. 352; (4) BD Pyxis MedStation 4000 Console, Catalog No. 309; (5) BD Pyxis MedStation 4000, Catalog No. 303; (6) BD Pyxis Anesthesia System 4000, Catalog No. 338
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
Labeling is insufficient regarding delays in access to medication. The firm is strengthening the label as an additional risk mitigation.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On January 8, 2025, the firm notified customers through an "Urgent Medical Device Correction" letter sent via email and mailed notices. In addition to explaining software issues affecting the device (covered in RES 96055), the letter also provided customers with a list of mitigations that address device downtime as a result of power failure, network failure, or hardware failure. The firm will be updating product labeling and product guides to further strengthen recommendations. On September 25, 2025, additional correction notices were distributed identifying additional affected devices. If you require further assistance contact the firm: Technical Support for technical questions: bd.com/self-service, 1-800-727-6102, Hours: 24 hours, 7 days a week; BD Recall Operations to return the customer response form: [email protected], Fax: (312) 949-0229
Quantity in commerce
193067
Distribution
Domestic distribution nationwide US. International distribution to North America, Latin America, EMEA, and Greater Asia.
Product codes and lots
(1)BD Pyxis MedStation ES, Catalog No. 323, UDI-DI 10885403512667; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327, UDI-DI 10885403477836; (3) BD Pyxis MedStation ES Tower, Catalog No. 352, UDI-DI 10885403512674; ALL serial numbers; software versions v1.10 and prior. (4) BD Pyxis MedStation 4000 Console, Catalog No. 309, All serial numbers and all software versions, UDI-DI: 10885403512636; (5) BD Pyxis MedStation 4000, Catalog No. 303, All serial numbers and all software versions, UDI-DI: 10885403512629 (HW), 10885403513787 (SW); (6) BD Pyxis Anesthesia System 4000, Catalog No. 338, All serial numbers and all software versions, UDI-DI: 10885403477829 (HW), 10885403513787 (SW)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1120-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1120-2025Class IIOngoingVoluntary: Firm initiated