Z-0976-2025Class II

BD Pyxis Med 4000 Auxiliary (AUX), REF 314

CareFusion 303, Inc / initiated 2023-02-14 / Open, Classified / root cause: manufacturing process

CareFusion 303, Inc. began a recall of BD Pyxis Med 4000 Auxiliary (AUX), REF 314 on . FDA classified it as Class II and gives the reason as "Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0976-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Pyxis Med 4000 Auxiliary (AUX), REF 314
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
BD issued an Urgent Medical Device Product Advisory to its consignees on 02/14/2023 via letter. The notice explained the problems with the device, health hazards, and requested the following actions be taken: "1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them. 3. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. 4. If you have specific questions about this letter, please contact 1-866-583-8783." The firm issued a second notice on 01/09/2024 with the same information.
Quantity in commerce
5 units
Distribution
Worldwide
Product codes and lots
UDI/DI 10885403512650, Serial Numbers: 16140589, 16194078, 16194079, 16194081, 16079165.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0976-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0971-2025Class IIOngoingVoluntary: Firm initiated
Z-0972-2025Class IIOngoingVoluntary: Firm initiated
Z-0973-2025Class IIOngoingVoluntary: Firm initiated
Z-0974-2025Class IIOngoingVoluntary: Firm initiated
Z-0975-2025Class IIOngoingVoluntary: Firm initiated
Z-0976-2025Class IIOngoingVoluntary: Firm initiated