Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized...
Draeger Medical Systems, Inc. began a recall of Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal... on . FDA classified it as Class II and gives the reason as "Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1066-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Draeger Medical Systems, Inc.
- Product
- Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Draeger notified consignees on about 01/14/2025 via Urgent - Medical Device Correction letter. Consignees were instructed to inspect the device to identify if the speaker is configured with the ICS in USB mode or Analog mode. For the speaker set to Analog mode there is an analog cable (aux cable) connected to the backside of the speaker and a light on frontside of the speaker display. If the light is green, there is no issue of speaker audio annunciation. If the light is red, the issue of speaker audio annunciation is present. If an issue with alarm annunciation occurs while operating, follow the steps included in the letter. Ensure all affected users or customers, if further distributed, are notified of the recall notification. They were instructed to complete and return the provided reply card.
- Quantity in commerce
- 1,013 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AS, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MA, NE, NY, OK, PA, SD, TN, TX, VA, WA and the countries of Argentina, Australia, Austria, Belgium, Botswana, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kenya, Malaysia, Mexico, Monaco, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Romania, Samoa, American, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkiye, United Kingdom.
- Product codes and lots
- Infinity CentralStation (ICS) Model Number: MS26800. UDI-DI Number: 04049098001878. Edifier Speaker Model Number: MS34036. UDI-DI Number: 04048675548027. Serial Numbers: 12946M00269, 12946M00314, 12946M00506, 12946M00519, 12946M00518, 12946M00553, 1OQ7WR-00103, 1OQ8CA-00249, 1OQ8CA-00254, 1OQ8CA-00229, 1OQ8CA-00221, 1OQ8CA-00282, 1OQ8CA-00314, 1OQ8CA-00193, 1OQ8CA-00200, 1OQ8CA-00194, 1OQ8CA-00181, 1OQ8CA-00178, 1OQ8CA-00248, 1OQ8CA-00125, 1OQ8CA-00071, 1OQ8CA-00032, 1OQ8CA-00171, 1OQ8CA-00164, 1OQ8CA-00149, 1OQ8CA-00069, 1OQ8CA-00057, 1OQ8CA-00153, 1OQ8CA-00150, 1OQ8CA-00148, 1OQ8CA-00165, 1OQ8CA-00155, 1OQ8CA-00097, 1OQ8CA-00113, 1OQ8CA-00106, 1OQ8CA-00075, 1OQ8CA-00052, 1OQ8CA-00067, 1OQ8CA-00128, 1OQ8CA-00132, 1OQ8CA-00268, 1OQ8CA-00198, 1OQ8CA-00255, 1OQ8CA-00151, 1OQ8CA-00342, 1OQ8CA-00391, 1OQ8CA-00321, 1OQ8CA-00402, 1OQ8CA-00395, 1OQ8CA-00297, 1OQ8CA-00240, 1OQ89M-00691, 1OQ89M-00690, 1OQ89M-00708, 1OQ89M-00693, 1OQ89M-00687, 1OQ89M-00707, 1OQ89M-00703, 1OQ89M-00679, 1OQ89M-00683, 1OQ89M-00692, 1OQ89M-00610, 1OQ89M-00613, 1OQ89M-00669, 1OQ89M-00635, 1OQ89M-00664, 1OQ89M-00697, P1OQ84F-00080, 1OQ89M-00684, 1OQ89M-00676, 1OQ89M-00695, 1OQ8CA-00086, 1OQ89M-00677, 1OQ8CA-00031, 1OQ8CA-00027, 1OQ8CA-00065, 1OQ8CA-00092, 1OQ8CA-00082, 1OQ89M-00602, 1OQ89M-00680, 1OQ89M-00675, 1OQ89M-00593, 1OQ89M-00686, 1OQ89M-00696, 1OQ89M-00681, 1OQ89M-00704, 1OQ89M-00674, 1OQ89M-00661, 1OQ89M-00663, P1OQ89M-00559, P1OQ89M-00540, 1OQ8CA-00134, 1OQ8CA-00095, 1OQ8CA-00043, 1OQ8CA-00074, 1OQ8CA-00762, 1OQ8CA-00688, 1OQ8CA-00670, 1OQ8CA-00637, 1OQ8CA-00633, 1OQ8E8-00037, 1OQ8E8-00053, 1OQ8CA-00773, 1OQ8CA-00683, 1OQ8CA-00500, 1OQ8CA-00476, 1OQ8CA-00449, 1OQ8CA-00420, 1OQ8CA-00460, 1OQ8CA-00463, 1OQ8CA-00507, 1OQ8CA-00350, 1OQ8CA-00442, 1OQ8CA-00433, 1OQ8CA-00447, 1OQ8CA-00601, 1OQ8CA-00604, 1OQ8CA-00695, 1OQ8CA-00513, 1OQ8CA-00575, 1OQ8CA-00577, 1OQ8CA-00591, 1OQ8CA-00197, 1OQ8CA-00516, 1OQ8CA-00541, 1OQ8CA-00542, 1OQ8CA-00568, 1OQ8CA-00499, 1OQ8CA-00498, P12051M00227, 1OQ8CA-00593, 12946M00532, 12946M00256, 12946M00278, 1OQ7WR-00203
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1066-2025 | Class II | Ongoing | Voluntary: Firm initiated |