Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1170-2025Class II

Brand Name: Sertera Biopsy Device Product Name: Sertera Biopsy Device Model/Catalog Number: Sertera-14...

Hologic, Inc. / initiated 2025-01-15 / Open, Classified / root cause: manufacturing process

Hologic, Inc began a recall of Brand Name: Sertera Biopsy Device Product Name: Sertera Biopsy Device Model/Catalog Number: Sertera-14 Software Version: N/A Product Description: Instrument... on . FDA classified it as Class II and gives the reason as "Devices presented a condition in which the inner needle separated from the hub and produced a longer throw of the inner needle or dislodgement". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1170-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Hologic, Inc.
Product
Brand Name: Sertera Biopsy Device Product Name: Sertera Biopsy Device Model/Catalog Number: Sertera-14 Software Version: N/A Product Description: Instrument, Biopsy Component: N/A
Product code
KNW Instrument, Biopsy
Reason for recall
Devices presented a condition in which the inner needle separated from the hub and produced a longer throw of the inner needle or dislodgement.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On January 15, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Our goal is to create as little disruption to your practice as possible. Below outlines the next steps for customers with affected product(s): " Please quarantine any product(s) from the impacted lots you currently have in your inventory. Impacted lots can be returned through IQVIA as part of the voluntary recall. Please note that you will be issued a credit on your account for the affected product(s). " Refer to the Customer Response Form Instructions to identify your product inventory that is subject to this action. " Confirm receipt of this communication by completing the online Customer Response Form. Please complete and submit the online Customer Response Form even if you no longer have the affected product in inventory. Hologic has partnered with IQVIA to conduct follow up communications should no response be received to this letter. " Hologic will issue a Return Goods Authorization upon receipt of the Customer Response Form and the return of all affected products to Hologic. " If you wish to reorder product, please place a new order via your preferred procurement method. We value your business and thank you for your cooperation. If you have any questions about this communication, please contact [email protected] or 1-877-371-4372.
Quantity in commerce
55,335 units
Distribution
Domestic: Nationwide Distribution
Product codes and lots
Model No- Sertera-14; UDI- 15420045504066; Lot Numbers: E23B23RL, E23B23RM, E23C07RD, E23C07RF, E23C31RC, E23D12RT, E23D12RU, E23D27RW, E23E03RJ, E23E12RT, E23E12RU, E23E29RL, E23E30RM, E23F02RG, E23F07RL, E23F07RM, E23F27R, E23F29RL, E23F29RM, E23F29RN, E23G12RK, E23G25R, E23G31RL, E23H02R, E23H02RA, E23H02RB, E23H28RA, E23H31RF, E23H31RG, E23K02RL, E23K10R, E23K10RA, E23K10RB, E23K31RU, E23L06RK, E23L10RB, E23L22RV, E23L28RB, E23L28RD, E23M21RV, E24A08RL, E24A08RM, E24A26RD, E24B01RF, E24B07RK, E24B22RC, E24B28R, E24B28RA, E24C07RA, E24C21R, E24D26RF, E24E02RD, E24E15RA, E24E22R, E24E22RA, E24E24RB.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1170-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1170-2025Class IIOngoingVoluntary: Firm initiated