Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient...
Fresenius Kabi USA, LLC began a recall of Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration... on . FDA classified it as Class I and gives the reason as "Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1019-2025
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Fresenius Kabi USA, LLC
- Product
- Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- Fresenius Kabi began notifying consignees on 01/10/2024 via emailed letter. Consignees were instructed to check if your device contains the affected software, review and become familiar with the temporary Risk Mitigation actions for each anomaly outlined in the letter and notify customers/personnel if the product was transferred or further distributed. Additionally, to facilitate the installation of the pump software, LVP SW 5.10, your facility must install the new Ivenix Infusion Management System (IMS) software version (5.2) on your IMS server and install the new software version (5.10) for the Ivenix Large Volume Pump (LVP). Please post this letter in areas that affected devices are stored and used such that users of this device are aware of the issues and recommended risk mitigation measures. Lastly, consignees were requested to complete and return the Customer Reply Form.
- Quantity in commerce
- 23 systems
- Distribution
- US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.
- Product codes and lots
- UDI-DI: 00811505030122. Software 5.9.2 and prior
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1019-2025 | Class I | Ongoing | Voluntary: Firm initiated |