Z-1115-2025Class II
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Siemens Healthcare Diagnostics Inc. / initiated 2024-01-03 / Completed / root cause: software
Beckman Coulter, Inc. began a recall of Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument on . FDA classified it as Class II and gives the reason as "A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results". As of , the recall status on record is Completed, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1115-2025
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- A field service engineer was dispatched to the customer site to perform the software update and to replace the PCB TSI-RSI board from 01/03/2024 through 12/26/2024.
- Quantity in commerce
- 61 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brazil, Croatia, France, Germany, Hungary, India, Ireland, Italy, New Zealand, Portugal, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland.
- Product codes and lots
- UDI/DI 15099590732103, serial numbers: 300208, 300201, 300184, 300185, 300169, 300225, 300175, 300172, 300221, 300209, 300211, 300212, 300213, 300214, 300215, 300210, 300174, 300226, 300204, 300216, 300217, 300170, 300220, 300176, 300188, 300199, 300200, 300177, 300222, 300224, 300173, 300181, 300202, 300191, 300223, 300171, 300186, 300190, 300203, 300178, 300195, 300189, 300198, 300227, 300228, 300168, 300194, 300205, 300206, 300207, 300197, 300218, 300179, 300182, 300192, 300219, 300193, 300183, 300180, 300187, 300196.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1115-2025 | Class II | Completed | Voluntary: Firm initiated |