Z-1115-2025Class II

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument

Siemens Healthcare Diagnostics Inc. / initiated 2024-01-03 / Completed / root cause: software

Beckman Coulter, Inc. began a recall of Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument on . FDA classified it as Class II and gives the reason as "A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results". As of , the recall status on record is Completed, and MDM tags the root cause as software.

Recall record

Recall number
Z-1115-2025
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.
Root cause tag
software
FDA root cause
Software design
Action
A field service engineer was dispatched to the customer site to perform the software update and to replace the PCB TSI-RSI board from 01/03/2024 through 12/26/2024.
Quantity in commerce
61 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brazil, Croatia, France, Germany, Hungary, India, Ireland, Italy, New Zealand, Portugal, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland.
Product codes and lots
UDI/DI 15099590732103, serial numbers: 300208, 300201, 300184, 300185, 300169, 300225, 300175, 300172, 300221, 300209, 300211, 300212, 300213, 300214, 300215, 300210, 300174, 300226, 300204, 300216, 300217, 300170, 300220, 300176, 300188, 300199, 300200, 300177, 300222, 300224, 300173, 300181, 300202, 300191, 300223, 300171, 300186, 300190, 300203, 300178, 300195, 300189, 300198, 300227, 300228, 300168, 300194, 300205, 300206, 300207, 300197, 300218, 300179, 300182, 300192, 300219, 300193, 300183, 300180, 300187, 300196.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1115-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1115-2025Class IICompletedVoluntary: Firm initiated