VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro...
Ventec Life Systems, Inc. began a recall of VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002 on . FDA classified it as Class I and gives the reason as "Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1196-2025
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ventec Life Systems, Inc.
- Product
- VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
- FDA root cause
- Component change control
- Action
- On 02/03/2025, recall notices were sent to Durable Medical Equipment Distributors asking them to do the following: Clinical Action: Follow internal procedures to transfer the patient to an alternative source of ventilation, if available. If no other ventilation method is available, you may continue to use the affected product adhering to the guidelines below, while you await device correction by firm Service personnel: 1. Use only when the patient is actively monitored to ensure timely response to a potential malfunction. 2. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 3. Use external monitors (e.g. SpO2) and set the alarm thresholds appropriately for the patient. 4. Connect the ventilator to an external monitoring system, wherever possible. Also, Durable Medical Equipment Distributors are asked to acknowledge they forwarded the recall notice, and that they contacted the users to which the devices were provided, advised them of the recall and action instructions within accompanying Recall Notification Letter including how to have the devices repaired. Complete and return the acknowledgement form via email to [email protected] For any questions contact the firm at at 844-MYVOCSN (844-698-6276), Monday through Friday between 8:00 a.m. - 5:00 p.m. PDT.
- Quantity in commerce
- 30
- Distribution
- US distribution in CA, NY, and MO.
- Product codes and lots
- REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421, 116700, 117933, 5038112, 5038619, 5038717, 5038953, 5039086, 5039091, 5039361, 5039499, 5039503, 5039541, 5039636, 5039938, 5040019, 5040640, 5040762, 5041060, 5041078, 5041534, 5042558, 117344, 118464, 118509, 5039387; PRT-01185-002/00850018761154/5037655