human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use
HUMAN MED AG began a recall of human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use on . FDA classified it as Class II and gives the reason as "The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1251-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Human Med AG
- Product
- human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
- Product code
- QPB System, Suction, Lipoplasty For Removal
- Reason for recall
- The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The recalling firm issued a Field Safety Notice to Distributors dated 10/18/2025 via email explaining the reason for the Field Safety Notice, the potential hazard, and informing the consignee to discontinue delivering the affected products to customers and users and to dispose of the product if still in their warehouse. A Field Safety Notice was enclosed to be issued to the consignee's users and customers who received the affected product from the Distributor. The User and Customer letter dated 10/17/2024 was similar to the Distributor letter except provided instructions to check the spray pattern as instructed in the user manuals prior to using the cannula. If the pattern was not in a flat, fan-shaped pattern, they are not to use the cannula and it should be sorted out and disposed of properly. A confirmation form was enclosed to confirm the cannula has been correctly disposed of and the form was to be sent to the distributor or to the email address provided.
- Quantity in commerce
- 775 devices
- Distribution
- US Nationwide distribution in the state of FL.
- Product codes and lots
- Lot number 24-20198, UDI (01)04260170881014(11)240611(17)270611(10)24-20198; and Lot number 23-19612, UDI (01)04260170881014(11)240109(17)270109(10)23-19612.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1251-2025 | Class II | Ongoing | Voluntary: Firm initiated |