Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or...
Augustine Temperature Management, LLC began a recall of Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia on . FDA classified it as Class II and gives the reason as "There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1199-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Augustine Temperature Management, LLC
- Product
- Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.
- Quantity in commerce
- 3 units
- Distribution
- US Nationwide distribution in the states of TX, AL, NV; AU, NZ.
- Product codes and lots
- UDI-DI UDI-DI 00855913001848, Serial Numbers: 1304384, 1304619, 1304620
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1197-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1198-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1199-2025 | Class II | Ongoing | Voluntary: Firm initiated |