DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software

Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. / initiated 2025-01-10 / Open, Classified / root cause: software

BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD. began a recall of DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software on . FDA classified it as Class II and gives the reason as "Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1233-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd.
Product
DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.
Root cause tag
software
FDA root cause
Software change control
Action
On 1/10/2025, recall notices were mailed to customers who were asked to do the following: 1) Replace and discard any reagent bottles that were on-board the analyzer at the time of any software upgrade. 2) Review Quality Control data and actions taken, for shifts out of allowed laboratory range which could have indicated potential deterioration in reagent stability. 3) Firm recommends sharing the content of this letter with your laboratory and/or Medical Director to determine if a review of previous patient test results should be conducted. 4) If specific software upgrade date is required, contact firm. 5) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. This issue will not occur with future software updates. If you have any questions regarding this notice, please contact the firm's Customer Support Center: Website: http://www.beckmancoulter.com Hotline: (800) 854-3633
Quantity in commerce
56
Distribution
US: MA, CA, SC, LA, OK, GA, AL, TX, OR, ID, CO, MN, NY, MS, PA, WA, IL, MI, PR. OUS: Lebanon, Poland, Ghana, Spain, South Africa, Slovakia, Ghana, Croatia, Seychelles, Australia, Republic of Korea
Product codes and lots
Software Versions: V1.4, V1.4.1, V1.4.2, V1.4.3 REF/UDI-DI/Serial Numbers: C63519/14987666545058/2024090170; C63520/14987666545065/2024070116, 2023070009, 2024010038, 2024010035, 2024050072, 2024040056, 2024030049, 2024020046, 2024070111, 2024030051, 2024030052, 2024010033, 2023090015, 2024060103, 2024060104, 2024060101, 2024080145, 2024020045, 2024020042, 2024020043, 2024020047, 2024060096, 2024040059, 2023100019, 2024030053, 2023070010, 2024070129, 2024070130, 2024020041, 2024090183, 2024010036, 2024010037, 2024070132, 2024070131, 2023100020, 2024050064, 2024050070, 2023090016, 2023110022, 2024070110, 2024060075, 2024060106, 2023110023, 2024010024, 2024010025, 2024050066, 2024010029, 2024060107, 2023080012, 2024040060, 2024050074, 2024040062, 2024010031, 2024030055, 2024080144