CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214
Smiths Medical ASD, Inc. began a recall of CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214 on . FDA classified it as Class II and gives the reason as "Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1405-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214
- Product code
- CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
- Reason for recall
- Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On February 14, 2025, customers were notified via Urgent Medical Device Correction letters. Consignees were instructed to locate and quarantine all affected product within their institution and destroy them. The recalling firm will contact customers who have returned completed response forms identifying affected products within their possession to discuss credit.
- Quantity in commerce
- 175,994 US; 45,223 OUS
- Distribution
- US Nationwide distribution. International distribution pending.
- Product codes and lots
- UDI-DI 30610586038444 Lots WW8214 022980-1-22 022404-1-17 021860-1-10 021110-1-7 020609-1-2 016288-1-1 022980-1-24 022404-1-18 021860-1-14 021110-1-8 020609-1-3 017774-1-1 022980-1-23 022404-1-19 021860-1-13 021110-1-9 020609-1-4 018139-1-1 022980-1-26 022404-1-20 021860-1-15 021110-1-10 020609-1-5 018184-1-1 022980-1-27 022404-1-21 021860-1-17 021110-1-11 020609-1-6 018252-1-1 022980-1-28 022404-1-22 021860-1-16 021110-1-12 020609-1-7 018458-1-1 022980-1-25 022404-1-23 021860-1-18 021110-1-13 020609-1-8 018457-1-1 022980-1-32 022404-1-24 021860-1-20 021110-1-14 020609-1-9 018614-1-1 022980-1-30 022404-1-25 021860-1-21 021110-1-16 020609-1-10 018641-1-1 022980-1-29 022404-1-26 021860-1-19 021110-1-15 020609-1-11 018789-1-1 022980-1-31 022404-2-2 021860-2-1 021110-1-17 020609-1-12 018789-2-1 022980-1-33 022404-1-27 021860-2-2 021110-1-18 020609-1-13 018789-3-1 023592-1-3 022404-2-1 021860-1-22 021110-1-19 020609-1-14 018789-4-1 023592-1-2 022404-2-3 021860-2-5 021110-1-20 020609-1-15 018936-1-1 023592-1-1 022404-2-4 021860-2-3 021110-1-21 020609-1-16 018936-2-1 023592-1-5 022404-2-5 021860-2-4 021110-1-22 020609-1-19 018936-3-1 023592-1-6 022404-2-6 021860-2-6 021110-1-23 020609-1-20 018936-4-1 023592-1-4 022404-2-7 021860-2-8 021110-1-24 020609-1-17 019179-1-1 023592-1-7 022404-2-8 021860-2-7 021110-1-25 020609-1-21 019254-1-1 023592-1-9 022404-2-9 021860-2-10 021110-1-26 020609-1-18 019323-1-1 023592-1-8 022404-2-10 021860-2-11 021110-1-27 020609-1-23 019383-1-1 023592-1-12 022404-2-11 021860-2-9 021110-1-28 020609-1-22 019409-1-1 023592-1-11 022404-2-12 021860-2-12 021110-1-29 020609-1-24 019410-1-1 023592-1-10 022404-2-13 021860-2-13 021110-1-30 020609-1-26 019549-1-1 023592-1-15 022980-1-3 021860-2-14 021110-1-33 020609-1-27 019613-1-1 023592-1-16 022980-1-1 021860-2-15 021110-1-31 020609-1-25 019671-1-1 023592-1-17 022980-1-2 021860-2-16 021110-1-34 020609-1-28 019747-1-1 023592-1-18 022980-1-4 021860-2-17 021110-1-35 020609-1-30 019782-1-1 023592-1-19 022980-1-5 021860-2-18 021110-1-32 020609-1-31 019810-1-1 023592-1-20 022980-1-6 022404-1-2 021110-1-36 020609-1-29 019920-1-1 9755313 022980-1-7 022404-1-1 021110-1-37 020609-1-34 019974-1-1 9755312 022980-1-8 022404-1-3 021110-1-38 020609-1-32 020000-1-1 9757103 022980-1-10 022404-1-4 021110-1-39 020609-1-33 020022-1-1 9757119 022980-1-11 022404-1-5 021110-1-40 020609-1-35 020022-2-1 9755311 022980-1-9 022404-1-6 021860-1-3 020609-1-36 020074-1-1 9759068 022980-1-12 022404-1-7 021860-1-4 020609-1-37 020126-1-1 9757614 022980-1-13 022404-1-8 021860-1-2 020609-1-38 020127-1-1 9761695 022980-1-15 022404-1-9 021860-1-1 020609-1-39 020159-1-1 9760404 022980-1-14 022404-1-10 021860-1-5 020609-1-40 020160-1-1 9762385 022980-1-16 022404-1-11 021860-1-7 021110-1-1 020227-1-1 9759947 022980-1-17 022404-1-12 021860-1-6 021110-1-2 020323-1-1 9759328 022980-1-18 022404-1-13 021860-1-11 021110-1-3 020346-1-1 9762393 022980-1-19 022404-1-14 021860-1-8 021110-1-4 020386-1-1 9763712 022980-1-20 022404-1-15 021860-1-9 021110-1-5 020435-1-1 9763703 022980-1-21 022404-1-16 021860-1-12 021110-1-6 020466-1-1 020609-1-1 020561-1-1 020551-1-1 020505-1-1 /
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1404-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1405-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.