Z-1405-2025Class II

CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214

Abbott Medical / initiated 2025-02-14 / Open, Classified / root cause: manufacturing process

Smiths Medical ASD, Inc. began a recall of CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214 on . FDA classified it as Class II and gives the reason as "Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1405-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214
Product code
CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Reason for recall
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On February 14, 2025, customers were notified via Urgent Medical Device Correction letters. Consignees were instructed to locate and quarantine all affected product within their institution and destroy them. The recalling firm will contact customers who have returned completed response forms identifying affected products within their possession to discuss credit.
Quantity in commerce
175,994 US; 45,223 OUS
Distribution
US Nationwide distribution. International distribution pending.
Product codes and lots
UDI-DI 30610586038444 Lots WW8214 022980-1-22 022404-1-17 021860-1-10 021110-1-7 020609-1-2 016288-1-1 022980-1-24 022404-1-18 021860-1-14 021110-1-8 020609-1-3 017774-1-1 022980-1-23 022404-1-19 021860-1-13 021110-1-9 020609-1-4 018139-1-1 022980-1-26 022404-1-20 021860-1-15 021110-1-10 020609-1-5 018184-1-1 022980-1-27 022404-1-21 021860-1-17 021110-1-11 020609-1-6 018252-1-1 022980-1-28 022404-1-22 021860-1-16 021110-1-12 020609-1-7 018458-1-1 022980-1-25 022404-1-23 021860-1-18 021110-1-13 020609-1-8 018457-1-1 022980-1-32 022404-1-24 021860-1-20 021110-1-14 020609-1-9 018614-1-1 022980-1-30 022404-1-25 021860-1-21 021110-1-16 020609-1-10 018641-1-1 022980-1-29 022404-1-26 021860-1-19 021110-1-15 020609-1-11 018789-1-1 022980-1-31 022404-2-2 021860-2-1 021110-1-17 020609-1-12 018789-2-1 022980-1-33 022404-1-27 021860-2-2 021110-1-18 020609-1-13 018789-3-1 023592-1-3 022404-2-1 021860-1-22 021110-1-19 020609-1-14 018789-4-1 023592-1-2 022404-2-3 021860-2-5 021110-1-20 020609-1-15 018936-1-1 023592-1-1 022404-2-4 021860-2-3 021110-1-21 020609-1-16 018936-2-1 023592-1-5 022404-2-5 021860-2-4 021110-1-22 020609-1-19 018936-3-1 023592-1-6 022404-2-6 021860-2-6 021110-1-23 020609-1-20 018936-4-1 023592-1-4 022404-2-7 021860-2-8 021110-1-24 020609-1-17 019179-1-1 023592-1-7 022404-2-8 021860-2-7 021110-1-25 020609-1-21 019254-1-1 023592-1-9 022404-2-9 021860-2-10 021110-1-26 020609-1-18 019323-1-1 023592-1-8 022404-2-10 021860-2-11 021110-1-27 020609-1-23 019383-1-1 023592-1-12 022404-2-11 021860-2-9 021110-1-28 020609-1-22 019409-1-1 023592-1-11 022404-2-12 021860-2-12 021110-1-29 020609-1-24 019410-1-1 023592-1-10 022404-2-13 021860-2-13 021110-1-30 020609-1-26 019549-1-1 023592-1-15 022980-1-3 021860-2-14 021110-1-33 020609-1-27 019613-1-1 023592-1-16 022980-1-1 021860-2-15 021110-1-31 020609-1-25 019671-1-1 023592-1-17 022980-1-2 021860-2-16 021110-1-34 020609-1-28 019747-1-1 023592-1-18 022980-1-4 021860-2-17 021110-1-35 020609-1-30 019782-1-1 023592-1-19 022980-1-5 021860-2-18 021110-1-32 020609-1-31 019810-1-1 023592-1-20 022980-1-6 022404-1-2 021110-1-36 020609-1-29 019920-1-1 9755313 022980-1-7 022404-1-1 021110-1-37 020609-1-34 019974-1-1 9755312 022980-1-8 022404-1-3 021110-1-38 020609-1-32 020000-1-1 9757103 022980-1-10 022404-1-4 021110-1-39 020609-1-33 020022-1-1 9757119 022980-1-11 022404-1-5 021110-1-40 020609-1-35 020022-2-1 9755311 022980-1-9 022404-1-6 021860-1-3 020609-1-36 020074-1-1 9759068 022980-1-12 022404-1-7 021860-1-4 020609-1-37 020126-1-1 9757614 022980-1-13 022404-1-8 021860-1-2 020609-1-38 020127-1-1 9761695 022980-1-15 022404-1-9 021860-1-1 020609-1-39 020159-1-1 9760404 022980-1-14 022404-1-10 021860-1-5 020609-1-40 020160-1-1 9762385 022980-1-16 022404-1-11 021860-1-7 021110-1-1 020227-1-1 9759947 022980-1-17 022404-1-12 021860-1-6 021110-1-2 020323-1-1 9759328 022980-1-18 022404-1-13 021860-1-11 021110-1-3 020346-1-1 9762393 022980-1-19 022404-1-14 021860-1-8 021110-1-4 020386-1-1 9763712 022980-1-20 022404-1-15 021860-1-9 021110-1-5 020435-1-1 9763703 022980-1-21 022404-1-16 021860-1-12 021110-1-6 020466-1-1 020609-1-1 020561-1-1 020551-1-1 020505-1-1 /

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1405-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1404-2025Class IIOngoingVoluntary: Firm initiated
Z-1405-2025Class IIOngoingVoluntary: Firm initiated