Alcon Custom Pak with Non-Latex Symbol, Pak: 12771-18, 18050-10, 18252-17, 12698-09, 15632-10, 3141-45...
Alcon Research LLC began a recall of Alcon Custom Pak with Non-Latex Symbol, Pak: 12771-18, 18050-10, 18252-17, 12698-09, 15632-10, 3141-45, 12630-07, 18925-05, 19265-05, 13894-16, 17550-08... on . FDA classified it as Class II and gives the reason as "Customer-designed sterile surgical procedure packs contain latex components as specified; but, the content label sealed within the transparent sterile pouch contains the symbol for a latex-free product. If users or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1293-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Alcon Laboratories
- Product
- Alcon Custom Pak with Non-Latex Symbol, Pak: 12771-18, 18050-10, 18252-17, 12698-09, 15632-10, 3141-45, 12630-07, 18925-05, 19265-05, 13894-16, 17550-08, 12236-18, 11854-15, 8043-29, 12466-17, 15599-17, 14764-09, 13194-16, 17069-09
- Product code
- LRO General Surgery Tray
- Reason for recall
- Customer-designed sterile surgical procedure packs contain latex components as specified; but, the content label sealed within the transparent sterile pouch contains the symbol for a latex-free product. If users or patients with latex sensitivity come into contact with latex, an allergic reaction may occur.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 2/10/2025, correction notices were mailed to customers who were asked to do the following: 1) Do not open packaging. Using the provided stickers, adhere one sticker to the outside of affected product near the label to identify it as containing latex. 2) Complete and return the response form electronically via: https://qrco.de/non-latex-symbol Should you have any questions or concerns about this matter, please feel free to call the firm at A 1-800-862-5266
- Quantity in commerce
- 2511
- Distribution
- US Nationwide distribution in the states of AK, AR, AZ, FL, MD, MN, NC, NE, OH, PA, TX, WA, WI.
- Product codes and lots
- UDI-DI: H5301ALCON1CPAK10. Pak/Lot(Expiration): 12771-18/JZ5190507(3/31/2025); 18050-10/JZ5200922(5/31/2025), JZ5202199(9/30/2025); 18252-17/JZ5200852(5/31/2025); 12698-09/JZ5201264(6/30/2025); 15632-10, JZ5202447(6/30/2025), JZ5202445(6/30/2025); 3141-45/ JZ5201303(7/31/2025), JZ5201305(7/31/2025); 12630-07/ JZ5201295(8/31/2025); 18925-05/JZ5194816(9/30/2025), JZ5194818(9/30/2025), JZ5196180(9/30/2025), JZ5196182(9/30/2025), JZ5200323(11/30/2025); 19265-05/JZ5201146(9/30/2025), JZ5201148(9/30/2025), JZ5202207(9/30/2025); 13894-16/JZ5201181(9/30/2025), JZ5202143(9/30/2025); 17550-08/JZ5198232(10/31/2025); 12236-18/JZ5201054(11/30/2025); 11854-15/JZ5201561(11/30/2025); 8043-29/JZ5202201(11/30/2025); 12466-17/JZ5205311(12/31/2025), JZ5205313(12/31/2025), JZ5205915(1/31/2026); 15599-17/JZ5205311(12/31/2025), JZ5205313(12/31/2025), JZ5205915(1/31/2026); 14764-09/JZ5197158(2/28/2026); 13194-16/JZ5197142(3/31/2026); 17069-09/JZ5201931(9/30/2026)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|